Getinge Disinfection Ab recalls

2 official recalls and safety alerts involving Getinge Disinfection Ab since 2025-05, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Getinge Disinfection Ab or the relevant agency.

  1. 2025-10-08
  2. US
    Drugs & medicalClass II

    Getinge Disinfection Ab recalls Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

    Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

    FDA · DeviceGetinge Disinfection AbWorldwide - US Nationwide distribution in the states of Colorado,…Official notice →
  3. 2025-05-14
  4. US
    Drugs & medicalClass II

    Getinge Disinfection Ab recalls Washer Disinfector Aquadis 56; Model Number: 56M

    The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the…

    FDA · DeviceGetinge Disinfection AbWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →