GE OEC Medical Systems recalls

2 official recalls and safety alerts involving GE OEC Medical Systems since 2023-08, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with GE OEC Medical Systems or the relevant agency.

  1. 2024-12-25
  2. US
    Drugs & medicalClass II

    GE OEC Medical Systems, Inc recalls OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to…

    Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

    FDA · DeviceGE OEC Medical Systems, IncU.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA,…Official notice →
  3. 2023-08-09
  4. US
    Drugs & medicalClass II

    GE OEC Medical Systems, Inc recalls OEC Flexiview 8800 Systems with 9-inch Image Intensifier

    Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which…

    FDA · DeviceGE OEC Medical Systems, IncWorldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →