Fort Defiance Industries recalls

2 official recalls and safety alerts involving Fort Defiance Industries since 2021-10, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Fort Defiance Industries or the relevant agency.

  1. 2026-08-26
  2. US
    Drugs & medicalClass II

    Fort Defiance Industries, LLC recalls Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

    The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

    FDA · DeviceFort Defiance Industries, LLCWorldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the…Official notice →
  3. 2021-10-27
  4. US
    Drugs & medicalClass II

    Fort Defiance Industries, LLC recalls P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641

    The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.

    FDA · DeviceFort Defiance Industries, LLCDistributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally…Official notice →