Erbe recalls

2 official recalls and safety alerts involving Erbe since 2026-04, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Erbe or the relevant agency.

  1. 2026-08-19
  2. US
    Drugs & medicalClass II

    Erbe USA Inc recalls ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

    A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive…

    FDA · DeviceErbe USA IncUS Nationwide distribution including in the states and territories of Alabama,…Official notice →
  3. 2026-04-01
  4. US
    Drugs & medicalClass I

    Erbe USA Inc recalls Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use

    Probes may rupture/burst during activation

    FDA · DeviceErbe USA IncUS Nationwide distribution, including Puerto Rico.Official notice →