Cytrellis Biosystems recalls
2 official recalls and safety alerts involving Cytrellis Biosystems since 2023-06, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Cytrellis Biosystems or the relevant agency.
- 2023-08-16
US
Cytrellis Biosystems, Inc. recalls ellacor System with Micro-Coring Technology-Indicated for use by medical professionals for the treatment of moderate and severe…
Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to…
- 2023-06-07
US
Cytrellis Biosystems, Inc. recalls Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the treatment of moderate and…
Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.