Civco Medical Instruments recalls
3 official recalls and safety alerts involving Civco Medical Instruments since 2021-12, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Civco Medical Instruments or the relevant agency.
- 2026-04-08
US
Civco Medical Instruments Co. Inc. recalls eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
- 2025-06-11
US
Civco Medical Instruments Co. Inc. recalls BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic…
During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
FDA · DeviceCivco Medical Instruments Co. Inc.OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and…Official notice →- 2021-12-29
US
Civco Medical Instruments Co. Inc. recalls VirtuTRAX Instrument Navigator, 14 Ga VirtuTRAX
There is potential that the VirtuTRAX device will not tighten onto the instrument to be used.
FDA · DeviceCivco Medical Instruments Co. Inc.Worldwide distribution - US Nationwide and the countries of Belgium, Canada,…Official notice →