Carestream Health recalls

3 official recalls and safety alerts involving Carestream Health since 2023-03, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Carestream Health or the relevant agency.

  1. 2023-12-13
  2. US
    Drugs & medicalClass II

    Carestream Health, Inc. recalls DRX Revolution Mobile X-Ray System

    Unexpected failure of electrical components within the CPI generator.

    FDA · DeviceCarestream Health, Inc.Worldwide Distribution: US (Nationwide Distribution); and OUS (Foreign):…Official notice →
  3. 2023-11-15
  4. US
    Drugs & medicalClass II

    Carestream Health, Inc. recalls The DRX-Revolution Mobile X-Ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. the system…

    There is a potential for unexpected failure of the electrical components within the Carestream Health Inc. generator installed in your DRX-Revolution Mobile X-ray System(s). This can lead to temporary and self…

    FDA · DeviceCarestream Health, Inc.US Nationwide. Global Distribution.Official notice →
  5. 2023-03-08
  6. US
    Drugs & medicalClass II

    Carestream Health, Inc. recalls DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to…

    After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially…

    FDA · DeviceCarestream Health, Inc.Worldwide distribution - US Nationwide distribution in the states of…Official notice →