- 2026-07-08
US
Accuray Incorporated recalls Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room…
FDA · DeviceAccuray IncorporatedWorldwide - US Nationwide and and the countries of Australia, Austria,…Official notice →- 2025-02-05
US
Accuray Incorporated recalls CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery…
Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.
- 2023-09-06
US
Accuray Incorporated recalls Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.
It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in…
FDA · DeviceAccuray IncorporatedWorldwide - US Nationwide distribution in the states of CA, GA, IL, LA, NJ,…Official notice →