Wishbone Medical, Inc. recalls WishFIX Growth Control Plating System, Model Number: TEFCS28-080US. Orthopedic device.
The products in this lot are incorrectly packed with a Stainless-Steel implant instead of a Titanium implant as the packaging and part number indicated.
Read the official notice →- Date
- 2024-05-22
- Country
- United States
- Brand
- Wishbone Medical
- Product
- WishFIX Growth Control Plating System, Model Number: TEFCS28-080US. Orthopedic device.
- Classification
- Class II
- Units
- US distribution to Florida, New Jersey and Connecticut.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Wishbone Medical recalls
- 2023-02-15
US
Wishbone Medical, Inc. recalls WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 10-702-080095-0 XS…
Firm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction.
FDA · DeviceWishbone Medical, Inc.US Nationwide distribution in the state of South Carolina.Official notice →
Other drugs & medical recalls in the United States
- 2024-05-22
US
QUIDELORTHO recalls VITROS Immunodiagnostic Products Free T3 Reagent Pack-IVD is for the quantitative measurement of free triiodothyronine (FT3) in human…
T3 Reagent Packs used with Free T3 Calibrators may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact…
FDA · DeviceQUIDELORTHOWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Baxter Healthcare Corporation recalls Spectrum IQ Infusion pump, Product Code 357009
There is a potential for cracks on the mount of the front panel of the device.
FDA · DeviceBaxter Healthcare CorporationUS Nationwide distribution in the states of Ohio, Wisconsin.Official notice →US
Karl Storz Endoscopy recalls LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000
Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then…
FDA · DeviceKarl Storz EndoscopyWorldwide distribution - US Nationwide in the states of OH, CA, FL, MA, TX,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Namic FLUID DELIVERY SET, REF 91300040
Products have a lack of sterility assurance.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the states of AL, FL, KY, MS, TX, WA, WI, NC,…Official notice →US
US
TriMed Inc. recalls TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref:…
Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
FDA · DeviceTriMed Inc.US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA,…Official notice →