Recall Atlas
Drugs & medicalClass I

Ventec Life Systems, Inc. recalls VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

Read the official notice →
Date
2026-10-28
Country
United States
Brand
Ventec Life Systems
Product
VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.
Classification
Class I
Units
Distribution US nationwide including Puerto Rico.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Ventec Life Systems recalls

  1. 2026-09-30
  2. Drugs & medicalClass II

    Ventec Life Systems, Inc. recalls REF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post Valve

    Labeling; Oxygen Cylinders do not contain required medical gas labeling

    FDA · DeviceVentec Life Systems, Inc.US Nationwide distribution in the states of FL, IA, OH, NE, TN.Official notice →
  3. 2026-07-01
  4. Drugs & medicalClass II

    Ventec Life Systems, Inc. recalls VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001

    Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of…

    FDA · DeviceVentec Life Systems, Inc.US Nationwide distribution in the state of UT.Official notice →
  5. 2026-05-13
  6. Drugs & medicalClass I

    Ventec Life Systems, Inc. recalls VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

    Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could…

    FDA · DeviceVentec Life Systems, Inc.US: OH, GA, CA, MOOfficial notice →
  7. 2025-03-05
  8. Drugs & medicalClass I

    Ventec Life Systems, Inc. recalls VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000,…

    Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on…

    FDA · DeviceVentec Life Systems, Inc.US distribution in CA, NY, and MO.Official notice →
  9. 2024-03-06
  10. Drugs & medicalClass I

    Ventec Life Systems, Inc. recalls VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN…

    The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions…

    FDA · DeviceVentec Life Systems, Inc.U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: JapanOfficial notice →
All 6 Ventec Life Systems recalls →

Other drugs & medical recalls in the United States

  1. 2026-10-28
  2. Drugs & medicalClass II

    Think Surgical, Inc. recalls Think Surgical Square Driver REF: 109126

    Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.

    FDA · DeviceThink Surgical, Inc.US Nationwide distribution in the states of NY and DE.Official notice →
  3. Drugs & medicalClass II

    Abbott recalls Amplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3. 9-ACP2-007-020 4. 9-ACP2-007-022 5.…

    Changes to Instructions for Use (IFU) for the device regarding risk of Pulmonary Artery Injury (PAI).

    FDA · DeviceAbbottWorldwide - US Nationwide distribution.Official notice →
  4. Drugs & medicalClass I

    Philips Ultrasound, LLC recalls Philips Lumify Diagnostic Ultrasound System

    Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital…

    FDA · DevicePhilips Ultrasound, LLCWorldwide - US Nationwide distribution including in the states of CA, MA, IL,…Official notice →
  5. Drugs & medicalClass I

    Par Health USA, LLC recalls Argatroban Injection, USP, 250 mg/2.5mL (100 mg/mL), 2.5 mL x 1 Single Dose Vial, Rx only, Manufactured for: Endo USA, Malvern,PA…

    Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.

    FDA · DrugPar Health USA, LLCNationwide within the United StatesOfficial notice →
See all drugs & medical recalls →