Tornier S.A.S. recalls Blueprint Software, BPUE001. Intended for use by surgeons as a pre surgical planner for shoulder replacement.
A software bug in Blueprint version 4.2.1 prevents users from accessing cases initiated in previous versions of the software.
Read the official notice →- Date
- 2025-05-14
- Country
- United States
- Brand
- Tornier
- Product
- Blueprint Software, BPUE001. Intended for use by surgeons as a pre surgical planner for shoulder replacement.
- Classification
- Class II
- Units
- Blueprint Version 4.2.1 is only currently available in the US. Domestic…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Tornier recalls
- 2026-04-15
US
Tornier, Inc recalls Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat,…
Affected products were incorrectly labeled as Tornier HRS Max products. These parts are compatible with the Tornier HRS system only, and are incompatible with the Tornier HRS Max system.
FDA · DeviceTornier, IncUS Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA,…Official notice →- 2026-02-11
US
Tornier, Inc recalls TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318
A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected…
- 2025-04-30
US
Tornier, Inc recalls stryker Latitude EV HUMERAL REVISION STEM TRIAL, L: 150mm, Size: LARGE, Side: RIGHT, REF 9030103; Prosthesis, Elbow, Semi-Constrained,…
A specific lot of Latitude Humeral Trial Stems has the incorrect direct part marking. " Laser Marked: 9030103, Humeral Stem Trial 150mm Large / Right, Yellow color indicator (incorrect) " Device Actually: 9030092,…
FDA · DeviceTornier, IncOUS (International) Distribution to countries: Sweden, Australia, United…Official notice →- 2025-04-16
US
Tornier S.A.S. recalls stryker Blueprint Mixed Reality Coracoid Clamp, Catalog Number MRUE202, packaged in stryker Blueprint Mixed Reality Instrument Kit,…
Stryker identified an elevated complaint rate related to difficulties using the Blueprint Mixed Reality Guidance Instrumentation system is multifactorial, but in each instance prohibits the system from meeting its…
FDA · DeviceTornier S.A.S.US: TX, MN, AK, PA, MA, TN, ID, and France, Canada, and IrelandOfficial notice →- 2025-01-22
US
Tornier S.A.S. recalls Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty.
The Pin Guide is being recalled because the metal tube may disassemble from the plastic handle. This can result in the Blueprint Mixed Reality headset being unable to detect the Pin Guide and potentially prolonging…
- 2024-10-09
US
Tornier, Inc recalls Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty
Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.
Other drugs & medical recalls in the United States
- 2025-05-21
US
L'Oreal USA recalls La Roche-Posay Laboratoire Dermatologique Effaclar Duo Dual Action Acne Treatment, 5.5% Benzoyl Peroxide Acne Medication, 40 mL (135 FL.…
Chemical Contamination: This recall has been initiated due to detected trace levels of benzene.
- 2025-05-14
US
Philips Medical Systems Nederland B.V. recalls IntelliSpace Cardiovascular software versions 6.x, 7.x, and 8.x; Model Number: 830089.
An issue was identified with the software when using the Echo Module of ISCV, the issue may result in missing or incomplete information in the report.
FDA · DevicePhilips Medical Systems Nederland B.V.Domestic: Nationwide Distribution International: Australia, Austria, Belgium,…Official notice →US
WANDERCRAFT SAS recalls Atalante X, Powered Exoskeleton. Atalante X is intended to perform ambulatory functions and mobility exercises, hands-free, in…
Potential for exoskeleton to lose lateral balance and cause patient to fall.
US
Galt Medical Corporation recalls GALT Guidewire REF SGW-051-07 Guidewires are intended for use in percutaneous procedures, to introduce and position catheters and other…
Due to a potential open seal in the sterile barrier packaging.
FDA · DeviceGalt Medical CorporationWorldwide - U.S. Nationwide distribution in the states of AL, FL, GA, IL, MA,…Official notice →US
Galt Medical Corporation recalls B Braun Interventional Coaxial Dilator REF: KIT-018-47, KIT-019-67, KIT-018-41, KIT-019-37 These Micro-Introducer Kits are intended to…
Due to a potential open seal in the sterile barrier packaging .
FDA · DeviceGalt Medical CorporationWorldwide - U.S. Nationwide distribution in the states of AL, FL, GA, IL, MA,…Official notice →US
Remote Diagnostic Technologies Ltd. recalls Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007
Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May…
FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide - US Nationwide distribution including in the states of AZ, VA, FL,…Official notice →