Drugs & medicalClass II

TMJ Solutions Inc recalls TMJ Bilateral Implants, REF: CHG020

Custom temporomandibular joint implant may have contained incorrect patient-specific components that led to a poor fit of the implant, leading to a right-sided open bite.

Read the official notice →
Date
2026-07-01
Country
United States
Brand
TMJ Solutions
Product
TMJ Bilateral Implants, REF: CHG020
Classification
Class II
Units
US Nationwide distribution in the state of MA.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More TMJ Solutions recalls

  1. 2025-11-26
  2. US
    Drugs & medicalClass I

    TMJ Solutions Inc recalls TMJ Bilateral Implant REF: CHG020 Sterile EO, Rx Only

    Patient-fitted temporomandibular joint implant contained discrepancy in screw hole position.

    FDA · DeviceTMJ Solutions IncUS: IL, OR, MS, TX, OK, UT, MA, MD, NY, PA, VA, FL, CA, MN, GA OUS: Canada,…Official notice →
  3. 2024-10-02
  4. US
    Drugs & medicalClass II

    TMJ Solutions Inc recalls Stryker TMJ Unilateral Implant, REF: GHG010, consisting of the following components: Ref: CHG010AM, Anatomical Bone Model, Accessory, REF:…

    Patient-fitted temporomandibular (TMJ) prothesis contained incorrect components, SterileEO

    FDA · DeviceTMJ Solutions IncUS Nationwide distribution in the state of MN.Official notice →
All 3 TMJ Solutions recalls →

Other drugs & medical recalls in the United States

  1. 2026-07-01
  2. US
    Drugs & medicalClass II

    Zimmer, Inc. recalls Off-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040241

    Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.

    FDA · DeviceZimmer, Inc.US Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    PFM MEDICAL INC. recalls Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310

    A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with…

    FDA · DevicePFM MEDICAL INC.US Nationwide distribution in the states of NY, MD.Official notice →
  4. US
    Drugs & medicalClass II

    Ventec Life Systems, Inc. recalls VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001

    Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of…

    FDA · DeviceVentec Life Systems, Inc.US Nationwide distribution in the state of UT.Official notice →
  5. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A

    Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and The countries of AE, AT, AU, BE,…Official notice →
  6. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Da Vinci ASSY, PSS, SP1098, Part Number: 380601

    Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled…

    FDA · DeviceIntuitive Surgical, Inc.Worldwide - US Nationwide distribution including in the states of CA, GA, OH,…Official notice →
  7. US
    Drugs & medicalClass I

    Medline Industries, LP recalls Medline Convenience Kits: 1) DYNJ905503F, Model Number: CV ANESTHESIA - ROOM SET UP; 2) DYNJ905503G, Model Number: CV ANESTHESIA - ROOM…

    The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified…

    FDA · DeviceMedline Industries, LPUS Nationwide.Official notice →
See all drugs & medical recalls →