TELEFLEX LLC recalls Iso-Gard Filter S, Tethered Cap, REF: a) 19211T, OUS only; b) 19212T, OUS only; Microbial medical gas filter, single-use
Incidents of device splitting or detaching during use
Read the official notice →- Date
- 2022-10-26
- Country
- United States
- Brand
- Teleflex
- Product
- Iso-Gard Filter S, Tethered Cap, REF: a) 19211T, OUS only; b) 19212T, OUS only; Microbial medical gas filter, single-use
- Classification
- Class I
- Units
- US nationwide distribution
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Teleflex recalls
- 2023-07-05
US
TELEFLEX LLC recalls Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170085
Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- 2023-06-07
US
TELEFLEX LLC recalls Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or…
Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step…
- 2023-05-24
US
TELEFLEX LLC recalls Teleflex (manufacturer: Vascular SOLUTIONS) D-Stat Flowable Hemostat, REF 4000; for sealing residual oozing of tissue tracts of femoral…
The product contains an expired diluent vial. If the impacted product is used, the sterility of the diluent component cannot be assured as it has passed beyond the studied shelf life of the product.
- 2023-02-01
US
TELEFLEX LLC recalls Endotracheal Tube labeled as the following: a. Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal; b. Endotracheal Tube…
Potential for pilot balloon non-inflation or cuff non-deflation.
FDA · DeviceTELEFLEX LLCWorldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →- 2023-01-04
US
TELEFLEX LLC recalls Pilling Wecksorb Cylindrical Sponges
Specific lots have not been irradiated to eliminate pyronema and therefore the sterility of the affected products may be compromised.
- 2022-11-23
US
Other drugs & medical recalls in the United States
- 2022-10-26
US
B. Braun Medical, Inc. recalls CARESITE SMALLBORE EXT SET T-PORT 5 IN., Product Code 470114, 100 units per carton. The A6 Access Device is a valve intended for the…
The extension set has a high probability of potential failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of…
FDA · DeviceB. Braun Medical, Inc.Distribution US nationwide, Canada and Saudi Arabia.Official notice →US
B. Braun Medical, Inc. recalls ULTRASITE US04TL Smallbore T-Port Extension Set, Product Code 474515, 100 units per carton. to provide needle-free access on I.V. and…
The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream…
FDA · DeviceB. Braun Medical, Inc.Distribution US nationwide, Canada and Saudi Arabia.Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN):…
Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result…
US
Fresenius Medical Care Holdings, Inc. recalls 0.9% Sodium Chloride Injection, USP, Each 100 mL contains: SODIUM CHLORIDE, USP - 900 mg, WATER FOR INJECTION, USP - qs,…
Lack of Assurance of Sterility: Leakage of 0.9% Sodium Chloride for Injection, 1L, 12pk Saline Solution.
US
LNK International, Inc. recalls NDC 0363-6171-09 Walgreens Sinus Pressure, Pain & Cough ACETAMINOPHEN/ PAIN RELIEVER DEXTROMETHORPHAN HBr/ COUGH SUPPRESSANT GUAIFENESIN/…
Boxes mislabeled to read "Decongestant Free", but the product contains Phenylephrine HCI 5mg
US
AuroMedics Pharma LLC recalls Acyclovir Sodium Injection, 500mg/10mL (50mg/mL), 10 mL Single Dose Vial, Rx only, Distributed by: AuroMedics Pharma LLC, 279…
Presence of Particulate Matter: Customer complaint for a dark red, brown and black particulate floating inside vial.