Synthes (USA) Products LLC recalls Radial Head Replacement System Implant is a one-piece, stemmed radial head replacement. The radial head is available in diameters of…
DePuy Synthes Radial Head Replacement System Contraindications added to Electronic Instructions for Use (e-IFU): Absolute contraindications include: Infection, sepsis, osteomyelitis.
Read the official notice →- Date
- 2022-05-18
- Country
- United States
- Brand
- Synthes (USA) Products
- Product
- Radial Head Replacement System Implant is a one-piece, stemmed radial head replacement. The radial head is available in diameters of 19,…
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Synthes (USA) Products recalls
- 2024-06-05
US
Synthes (USA) Products LLC recalls STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during…
Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay
FDA · DeviceSynthes (USA) Products LLCWorldwide distribution - US Nationwide and the country of Canada.Official notice →- 2024-03-13
US
Synthes (USA) Products LLC recalls Opal, Intervertebral Cage, 10x32mm, h 11mm, Revolvable-Indicated for degenerative spine disease Part Number: 08.803.231S
Products not sterilized, sterility cannot be confirmed
FDA · DeviceSynthes (USA) Products LLCWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →- 2023-03-22
US
Synthes (USA) Products LLC recalls TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur…
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length
FDA · DeviceSynthes (USA) Products LLCWorldwide - US Nationwide distribution in the states of MS, OR, TX and the…Official notice →- 2022-10-26
US
Synthes (USA) Products LLC recalls SPEED 9X10X10mm Continuous Compression Implant, Product No. SE-0910
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide…
- 2022-10-12
US
Synthes (USA) Products LLC recalls Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001
Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate…
- 2022-02-16
US
Synthes (USA) Products LLC recalls Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the…
Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not…
Other drugs & medical recalls in the United States
- 2022-05-25
US
Plastikon Healthcare LLC recalls Magnesium Hydroxide 1200mg, Aluminum Hydroxide 1200mg, Simethicone 120mg, 30 mL cup, packaged in 10 cups per tray, 10 trays per carton,…
Microbial Contamination of Non-Sterile Products
FDA · DrugPlastikon Healthcare LLCDistributed in Indiana for further distribution in the USA.Official notice →- 2022-05-18
US
Shanghai United Imaging Healthcare Co., Ltd. recalls Digital Medical X-ray Imaging System, Model: uDR 596i, uDR 592h
X-ray imaging system positioning image and protocol label is reversed for Flexion and Extension on C-Spine and L-Spine, this will cause the image to be incorrectly labeled and may cause the patient to have a repeat…
FDA · DeviceShanghai United Imaging Healthcare Co., Ltd.US Nationwide distribution in the state of TX.Official notice →US
Daavlin Distributing Company recalls 1 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v. Models: [1 Ser…
Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.
FDA · DeviceDaavlin Distributing CompanyWorldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →US
Microtek Medical Inc. recalls TMJ Arthroscopy Drape, 85''x 96''
Latex content in product was labeled incorrectly as latex free.
US
Certified Safety Mfg Inc recalls The following first aid kits and cabinets containing components Medi-First XS Non-Aspirin, Product #80450, 2 tablets/packet, 25…
First aid kit and cabinet components are being recalled by the supplier due to cGMP deviations at the manufacturer.
FDA · DeviceCertified Safety Mfg IncDistribution was made to AL, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MD,…Official notice →US
bioMerieux, Inc. recalls VITEK 2 Systems and VITEK 2 with MYLA.
Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected interpretation. This can potentially lead to incorrect final…
FDA · DevicebioMerieux, Inc.U.S. distribution of the FSCA to NC. OUS distribution of the FSCA to…Official notice →