Stryker Neurovascular recalls Trevo ProVue, Catalog: 90184

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable…

Read the official notice →
Date
2024-06-05
Country
United States
Brand
Stryker Neurovascular
Product
Trevo ProVue, Catalog: 90184
Classification
Class II
Units
Worldwide - US Nationwide distribution including in the states of MN, PA, IL,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Stryker Neurovascular recalls

  1. 2026-07-01
  2. US
    Drugs & medicalClass II

    Stryker Neurovascular recalls INZONE DETACHMENT SYSTEM, REF: M00345100950

    Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended,…

    FDA · DeviceStryker NeurovascularWorldwide - US Nationwide distribution including in the states of PA, AZ, FL,…Official notice →
  3. 2024-10-16
  4. US
    Drugs & medicalClass II

    Stryker Neurovascular recalls The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end…

    Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed…

    FDA · DeviceStryker NeurovascularU.S. Nationwide distribution in the states of CA and VA.Official notice →
  5. 2024-06-12
  6. US
    Drugs & medicalClass II

    Stryker Neurovascular recalls The Synchro2 Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The…

    Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer…

    FDA · DeviceStryker NeurovascularWorldwide - US Nationwide distribution including in the states of TX, NY, MD,…Official notice →
  7. 2024-05-08
  8. US
    Drugs & medicalClass II

    Stryker Neurovascular recalls The Distal Access Catheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker on the distal end and a…

    Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification…

    FDA · DeviceStryker NeurovascularUS Nationwide distribution in the states of TX, FL, NJ, CO, LA, NC, AL, PA,…Official notice →
  9. 2022-07-27
  10. US
    Drugs & medicalClass II

    Stryker Neurovascular recalls Trevo Trak 21 Microcatheter, REF: 90338. TREVO NXT 3X32+TRAK21 2-PACK - CE, REF: 91412; TREVO NXT 4X28+TRAK21 2-PACK - CE, REF: 91413;…

    Microcatheter product label on the carton is missing the microcatheter product drawing/illustration.

    FDA · DeviceStryker NeurovascularWorldwide distribution - US Nationwide distribution in the states of CA, OR,…Official notice →
All 6 Stryker Neurovascular recalls →

Other drugs & medical recalls in the United States

  1. 2024-06-05
  2. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Medline medical procedure kits, labeled as follows: a) PARACERVICAL PUD TRAY, REF DYNDH1832; b) AHS PARACERVICAL/PUDENDAL…

    Trumpet Needle Guide ring can detach when excessive pressure is applied.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Synthes (USA) Products LLC recalls STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during…

    Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay

    FDA · DeviceSynthes (USA) Products LLCWorldwide distribution - US Nationwide and the country of Canada.Official notice →
  4. US
    Drugs & medicalClass II

    Stryker Orthopaedics recalls Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1…

    Application software intended to be deployed on Mako 3.0, part number 209999, and Mako 3.1, part number 219999, observed an increase in the Software Error #3 (SE3) error code when a Mako System shutdown or a Mako…

    FDA · DeviceStryker OrthopaedicsWorldwide - US Nationwide and the countries of Australia, Austria, Belgium,…Official notice →
  5. US
    Drugs & medicalClass II

    Smith & Nephew Inc recalls smith&nephew OXINIUM Femoral Head 12/14 Taper, REF 71343204, 32 MM OD +4, Sterile; and smith&nephew, 12/14 Taper, REF 71343604, 36 MM M / +4

    Packaging contained the incorrect size of femoral head from what was displayed on the label.

    FDA · DeviceSmith & Nephew IncInternational distribution to the countries of Australia, Germany, Japan,…Official notice →
  6. US
    Drugs & medicalClass II

    Lucid Diagnostics, Inc. recalls Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter, 10 pores/inch) Model/Catalog Number: EC25T10VI Software Version: N/A…

    The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper…

    FDA · DeviceLucid Diagnostics, Inc.US Nationwide distribution in the states of MN, MD.Official notice →
  7. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column,…

    Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

    FDA · DeviceFUJIFILM Healthcare Americas CorporationUS Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.Official notice →