Stryker Corporation recalls Stryker, StrykeFlow REF: 0250070520 / 2, Disposable Suction/Irrigator with Disposable Tip
Due to complaints their is the potential that irrigation solution may leaked into the handpiece and battery pack causing it to malfunction.
Read the official notice →- Date
- 2025-07-30
- Country
- United States
- Brand
- Stryker
- Product
- Stryker, StrykeFlow REF: 0250070520 / 2, Disposable Suction/Irrigator with Disposable Tip
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution including in the states of Alabama,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Stryker recalls
- 2026-02-25
US
Stryker Corporation recalls MOLLI 2 System System includes: MM1000 (Pack of 10) - MOLLI Introducer (8cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195028 MS-00009…
Due to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during target tissue removal procedures.
FDA · DeviceStryker CorporationWorldwide - U.S. Nationwide distribution including in the states of AL, AK,…Official notice →- 2025-12-03
US
Stryker Corporation recalls NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita;…
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
FDA · DeviceStryker CorporationWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2025-08-13
US
Stryker Corporation recalls Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK /…
Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral…
FDA · DeviceStryker CorporationDomestic US distribution nationwide. International distribution to Australia…Official notice →- 2025-06-18
US
Stryker GmbH recalls Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S.
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.
FDA · DeviceStryker GmbHWorldwide - US Nationwide distribution in the states of California and…Official notice →US
Stryker GmbH recalls Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.
Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.
FDA · DeviceStryker GmbHWorldwide - US Nationwide distribution in the states of AL, FL, NJ, OR and the…Official notice →- 2025-04-23
US
Stryker GmbH recalls Guide Wire with Ruler Tube 3x800 mm DIA; Catalog number/REF: 2351-3080S;
The metal ring at the end of the Guide Wire with Ruler Tube may detach from the main body of the instrument during use in surgery.
FDA · DeviceStryker GmbHUS Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NY, OH,…Official notice →
Other drugs & medical recalls in the United States
- 2025-08-06
US
Pfizer Inc. recalls Bicillin L-A (penicllin G benzathine injectable suspension), 1,200,000 units per 2 mL, 2 mL-vial, Rx Only, Distributed by Pfizer Inc., New…
CGMP Deviations; particulates identified during visual inspection
US
FDC Limited recalls Timolol Maleate Ophthalmic Solution USP, 0.5%, Sterile, 5mL bottles, Rx only, Manufactured by: FDC Limited, Waluj, Aurangabad,…
Defective Container: spike of the cap becomes lodged in the nozzle of the product bottle.
US
Lupin Pharmaceuticals Inc. recalls Amlodipine and Benazepril HCl Capsules USP 2.5 mg/10 mg, 100 Capsules bottle, Rx Only, Manufactured for: Lupin Pharmaceuticals, Inc.,…
Labeling: Incorrect or Missing Lot and/or Exp Date: released with wrong expiry date as Feb.2027 instead of Jan.2027
- 2025-07-30
US
Medtronic, Inc. recalls Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application,…
In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's…
US
CooperVision, Inc. recalls Biofinity XR Toric Contact Lens
Lens blisters may have an incomplete or leaking seal which may render them unsterile.
FDA · DeviceCooperVision, Inc.Worldwide distribution - US Nationwide and the countries of Brazil, Canada,…Official notice →US
Johnson & Johnson Vision Care, Inc. recalls ACUVUE¿ OASYS MAX 1-Day MULTIFOCAL
Due to defects (bubbles/voids) identified during standard finished goods testing
FDA · DeviceJohnson & Johnson Vision Care, Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, AZ,…Official notice →