St. Jude Medical recalls Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System…
Due to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.
Read the official notice →- Date
- 2025-01-29
- Country
- United States
- Brand
- St. Jude Medical
- Product
- Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System…
- Classification
- Class II
- Units
- U.S.: CA, CT, GA, IL, MI, MN, NC, NJ, NY, OR, PA, TX, and WA. O.U.S.: N/A
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More St. Jude Medical recalls
- 2025-04-02
US
St. Jude Medical recalls CardioMEMS Heart Failure (HF) System that consist of: -CardioMEMS Hospital System Model CM3100 -CardioMEMS Backend Web Application Model…
Due to issues associated with data migration of patient information and results to a Cloud service, the firm has received complaint associated with some patient profile information being duplicated while other patient…
FDA · DeviceSt. Jude MedicalU.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA,…Official notice →- 2025-01-29
US
St. Jude Medical recalls CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Hosptial Electronics System (PES) REF CM3000 Merlin.net Patient Care…
As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
FDA · DeviceSt. Jude MedicalWorldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →- 2024-07-24
US
St. Jude Medical recalls St. Jude Medical Agilis NxT Steerable Introducer, REF 408309
One lot of product has dilators that are too short and will not extend outside the introducer sheath.
- 2023-11-15
US
St. Jude Medical recalls CardioMEMS Patient Electronic System (PES), Model Number CM1100. Used for wirelessly measuring and monitoring pulmonary artery pressure…
Potential for damaged and frayed power connector plug with repeated bending or manipulation beyond 90 degrees.
FDA · DeviceSt. Jude MedicalWorldwide distribution - US Nationwide and the countries of ARGENTINA,…Official notice →- 2023-03-15
US
St. Jude Medical recalls CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA)…
Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of…
FDA · DeviceSt. Jude MedicalWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →US
St. Jude Medical recalls CardioMEMS HF System Hospital Electronics System (HES), Model number CM3000, The CardioMEMS HF System provides pulmonary artery (PA)…
Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits…
FDA · DeviceSt. Jude MedicalWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
Other drugs & medical recalls in the United States
- 2025-01-29
US
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mm,35 cm,…
The probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution and International: Algeria, Armenia, Australia,…Official notice →US
MICROVENTION INC. recalls Microvention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN, Sterile R, Rx Only, Lot# 0000629846
Sterility assurance; microcatheter inner packaging may contain incomplete seal.
US
St. Jude Medical recalls CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Hosptial Electronics System (PES) REF CM3000 Merlin.net Patient Care…
As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
FDA · DeviceSt. Jude MedicalWorldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →US
Medtronic Perfusion Systems recalls DLP Pediatric One-Piece Arterial Cannulae, Model Numbers 77008 (8 Fr) and 77014 (14 Fr)
Incorrect labeling for seven manufactured lots of certain models of Arterial Cannulae.
US
O&M HALYARD, INC. recalls Brand Name: HALYARD Product Name: Laparotomy Pack Model/Catalog Number: 88231 Software Version: N/A Product Description: Laparotomy Pack…
Elevated residues of anti-block/slip additive used in manufacturing found within sterile packaging.
FDA · DeviceO&M HALYARD, INC.US Nationwide distribution in the states of CO, FL, GA, HI, IA, IL, IN, KS,…Official notice →US
Stryker Corporation recalls stryker 3.0MM Neuro Match Head, REF 5820-107-430, used to cut bone and bone cement during neurosurgery
There is potential the device inside the package may not match the product information on the outer label (e.g., label states 5820-107-430 but package contains a 5820-107-530 device or vice versa).
FDA · DeviceStryker CorporationWorldwide distribution - US Nationwide and the countries of UK, Netherlands,…Official notice →