SPINEART SA recalls JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S;…
JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could…
Read the official notice →- Date
- 2023-05-03
- Country
- United States
- Brand
- Spineart
- Product
- JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S;…
- Classification
- Class II
- Units
- US Nationwide distribution in the states of NH, TX, CA, VA, IL, RI, NC, AZ.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Spineart recalls
- 2025-08-20
US
SPINEART SA recalls Brand Name: PERLA ¿ TL MIS Product Name: CANNULATED FENESTRATED POLYAXIAL SCREW Model/Catalog Number: MPF-PS 55 45-S Software Version: Not…
Cannulated fenestrated polyaxial screws may be incorrectly labeled
FDA · DeviceSPINEART SAUS Nationwide distribution in the states of Florida , Kentucky, California.Official notice →- 2024-05-15
US
SPINEART SA recalls PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55…
Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but…
FDA · DeviceSPINEART SAU.S. Nationwide distribution in the states of CA, and KY. Not provided due to…Official notice →
Other drugs & medical recalls in the United States
- 2023-05-03
US
ROi CPS LLC recalls Sterile surgical convenience kit: regard Item Number 800706005, LD0091E - OH C-Section Pack.
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
FDA · DeviceROi CPS LLCUS Nationwide distribution in the states of FL and MO. There was no…Official notice →US
Aesculap Implant Systems LLC recalls Caspar Cervical Retractor (CCR) Basket ME754 - Lid Only and ME764 Casper Cervical Retractor (CCR) Basket Full System - Lid Only. A…
Mislabeled: Product is marked with GTIN # 04046955299592 however; the correct GTIN # is 04046955299607.
FDA · DeviceAesculap Implant Systems LLCUS Nationwide distribution in the states of CA, OH, NE, NH, FL, PA, TX.Official notice →US
Hamilton Medical AG recalls HAMILTON-C6, REF: 160021
Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an…
FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA,…Official notice →US
Richard Wolf GmbH recalls FLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a…
FDA · DeviceRichard Wolf GmbHWorldwide distribution - US Nationwide and the countries of United Arab…Official notice →US
Carefusion 2200 Inc recalls (1) Carefusion V. Mueller Graves Vaginal Speculum, Catalog #GL11, Medium Blade, 4 x 1 3/8" (10.2 X 3.5cm); and (2) Carefusion V. Mueller…
A portion of the two products were incorrectly marked (etched) and packaged resulting in a product mismatch. The labels and laser etchings were switched on the two products.
FDA · DeviceCarefusion 2200 IncDistribution was made to AL, AR, AZ, CA, CO, CT, FL, HI, IA, IL, IN, KY, LA,…Official notice →US
Sun Pharmaceutical Industries Ltd. recalls Fyremadel (ganirelix acetate) injection, 250 mcg/0.5 mL, One Single Dose Sterile Prefilled Syringe 0.5 mL, For Subcutaneous…
Presence of Particulate Matter: A piece of glass was found in a prefilled syringe.