Smiths Medical ASD Inc. recalls CADD-Solis Ambulatory Infusion Pumps, Model 2100. Intended for the following uses: Indicated for the following uses: Intra-arterial,…
CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Upstream Occlusion, 2. Stop and Power Keys Unresponsive, 3. Manual Mode Air Detector, 4. Single Bubble Air…
Read the official notice →- Date
- 2024-08-14
- Country
- United States
- Brand
- Smiths Medical ASD
- Product
- CADD-Solis Ambulatory Infusion Pumps, Model 2100. Intended for the following uses: Indicated for the following uses: Intra-arterial,…
- Classification
- Class I
- Units
- Worldwide
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Smiths Medical ASD recalls
- 2024-09-18
US
Smiths Medical ASD Inc. recalls Pneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NXX
There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.
- 2024-09-04
US
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM STANDARD FLEXTEND PLUS , Product Code/List…
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
- 2024-08-14
US
Smiths Medical ASD Inc. recalls CADD-Solis VIP ambulatory Infusion Pumps, Model 2120 ( 21-2120 , 21-2125 , and 21-2127 ) Indicated for the following uses:…
The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
- 2024-04-24
US
Smiths Medical ASD Inc. recalls Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-51, d) 4000-0105-249, e)…
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead…
- 2024-03-27
US
Smiths Medical ASD Inc. recalls smiths medical Pneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NUS
When a paraPAC plus ventilator is switched to the operating mode of VENTILATE, the ventilator may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This…
FDA · DeviceSmiths Medical ASD Inc.Worldwide distribution: US (nationwide) and OUS countries of: Algeria,…Official notice →- 2024-02-07
US
Smiths Medical ASD Inc. recalls smiths medical Medfusion Model 4000 syringe pump
Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2.…
FDA · DeviceSmiths Medical ASD Inc.US nationwide, Canada, Great Britain, Bermuda, Australia, Germany, Singapore,…Official notice →
Other drugs & medical recalls in the United States
- 2024-08-14
US
Covidien recalls Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR…
Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than…
FDA · DeviceCovidienUS: KY, OH, GA, MN, IN,UT, IL, NJ, NY, PR, LA, MD, MO, TN, MA, ME, CA, SC, AR,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline procedure kits labeled as: 1) NEPHROLOGY PACK, Pack Number DYNJ59089; 2) NERVE BLOCK TRAY, Pack Number DYNJRA1221A
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide Distribution: US (nationwide) and OUS countries of: AE, AI, BB, CA,…Official notice →US
Smiths Medical ASD Inc. recalls CADD-Solis VIP ambulatory Infusion Pumps, Model 2120 ( 21-2120 , 21-2125 , and 21-2127 ) Indicated for the following uses:…
The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
US
Granules Pharmaceuticals Inc. recalls Gabapentin Tablets, USP, 600 mg, 500-count bottles, Rx only, Manufactured by: Granules India Limited Hyderabad-500 081, India,…
Presence of Foreign Tablets; 3 fused tablets of Metformin ER 500 mg were found in bottle of Gabapentin Tablets
US
A-S Medication Solutions LLC recalls Acetaminophen Extra Strength Tablets, 500 mg each, 100 count bottles, Packaged By: A-S Medication Solutions, Libertyville, IL…
Labeling: Label Mix-Up: Some bottles of Acetaminophen Extra Strength 500 mg tablets were incorrectly labeled with the drug facts label for Aspirin 81 mg tablets.
US