Siemens Medical Solutions USA, Inc recalls Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version…
Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.
Read the official notice →- Date
- 2026-02-04
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version VB10D-SP02
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of AK, CA, IL, MN, NY,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2026-04-22
US
Siemens Medical Solutions USA, Inc recalls ARTIS Pheno VE30A and VE40A, Model 10849000
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the…
- 2026-04-15
US
Siemens Medical Solutions USA, Inc recalls ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels…
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution.Official notice →- 2026-02-11
US
Siemens Medical Solutions USA, Inc recalls NAEOTOM Alpha.Pro Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide distribution - US Nationwide and the countries of Canada, Costa…Official notice →- 2026-01-14
US
Siemens Medical Solutions USA, Inc recalls LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471
A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ…
- 2025-12-10
US
Siemens Medical Solutions USA, Inc recalls MAMMOMAT Inspiration;
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution in the states of AR, CA, NY and the…Official notice →- 2025-11-19
US
Siemens Medical Solutions USA, Inc recalls MAGNETOM Cima.X Upgrade. Model Number: 11689304.
There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting…
FDA · DeviceSiemens Medical Solutions USA, IncUS distribution to California. International distribution to Australia,…Official notice →
Other drugs & medical recalls in the United States
- 2026-02-04
US
NIPRO Technical Services, Inc. recalls Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard…
due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.
FDA · DeviceNIPRO Technical Services, Inc.U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL,…Official notice →US
GOLD STAR DISTRIBUTION INC recalls Feminine hygiene products, labeled as: KOTEX FREEDOM MAXI NO WINGS 48CT, ALWAYS THIN LINERS UNSCENTED 12 CT, ALWAYS THIN LINERS…
Potential exposure of rodents and rodent activity in the distribution center.
FDA · DeviceGOLD STAR DISTRIBUTION INCUS Nationwide distribution in the states of Minnesota.Official notice →US
Canon Medical System, USA, INC. recalls Alphenix INFX-8000F, interventional fluoroscopic x-ray system
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result,…
FDA · DeviceCanon Medical System, USA, INC.Worldwide distribution - US Nationwide and the country of Dominican Republic.Official notice →US
Aju Pharm Co., Ltd. recalls AlternatiV+ Screw-In Anchor
Reports of device failure, including anchor breakage, anchor bending, anchor pull out, suture breakage, and secondary packaging breakage.
FDA · DeviceAju Pharm Co., Ltd.US Nationwide distribution in the states of CA & PR.Official notice →US
Cerapedics, Inc. recalls PearMatrix P-15 Peptide Enhanced Bone Graft Bone Graft, 1.0cc REF: 730-010
Incorrect expiration date
FDA · DeviceCerapedics, Inc.US Nationwide distribution in the states of CA, CO, CT, FL, GA, ID, IL, KS,…Official notice →US
Elekta, Inc. recalls Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD Software Version:…
Due to a manufacturing issue and electrical grounding of systems.