Drugs & medicalClass II

Siemens Medical Solutions USA, Inc recalls Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care…

Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of…

Read the official notice →
Date
2023-08-02
Country
United States
Brand
Siemens Medical Solutions
Product
Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station,…
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of FL, GA, KS, MN, NC, NY…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Medical Solutions recalls

  1. 2023-09-20
  2. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc. recalls ACUSON Redwood 2.0 ultrasound systems, REF: 11503314

    If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6…

    FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide - US Nationwide distribution including in the states of IL, WI, CA,…Official notice →
  3. 2023-08-16
  4. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material…

    Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable…

    FDA · DeviceSiemens Medical Solutions USA, IncWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →
  5. 2023-08-09
  6. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc. recalls The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor…

    During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the…

    FDA · DeviceSiemens Medical Solutions USA, Inc.Distribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no…Official notice →
  7. 2023-05-31
  8. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Luminos Agile Max (VE10, VF10, VF11)

    Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point,…

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide - Worldwide DistributionOfficial notice →
  9. 2023-05-10
  10. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM)- Angiography systems developed for single…

    If, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range…

    FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution.Official notice →
  11. 2023-05-03
  12. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ARTIS one with a hybrid cable- An angiography system developed for diagnostic imaging and interventional procedures. Material…

    Examination room monitor connected with the hybrid cable may have a display issue and no image function is available

    FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution in the states of FL, SD, PR and the…Official notice →
All 70 Siemens Medical Solutions recalls →

Other drugs & medical recalls in the United States

  1. 2023-08-09
  2. US
    Drugs & medicalClass II

    The Ritedose Corporation recalls Cromolyn Sodium Oral Solution (Concentrate) 100 mg/5 mL 96 Plastic Ampules per carton, Rx Only, For Oral Use Only - Not For Inhalation…

    CGMP deviation: product released prior to supplier approval.

    FDA · DrugThe Ritedose CorporationNationwide in the USOfficial notice →
  3. 2023-08-02
  4. US
    Drugs & medicalClass II

    Qiagen GmbH recalls QIAstat-Dx Respiratory SARS-CoV-2 Panel -US IVD qualitative test intended for analyzing nasopharyngeal swab (NPS) samples from patients…

    Identified a decreased performance reliability rate, Run abortions could cause delayed diagnosis since sample testing would have to be repeated

    FDA · DeviceQiagen GmbHDC, AL, CA, FL, GA, ME, NC, NY, OR, TX Foreign: DE, AE, AT, BE, BG, CH, CL,…Official notice →
  5. US
    Drugs & medicalClass II

    Davol, Inc. recalls Ventralight ST w Echo PS 4.5" Circle A low profile, bioresorbable, coated, permanent mesh, with a pre-attached removable positioning…

    Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration,…

    FDA · DeviceDavol, Inc.Worldwide distribution - US Nationwide and the countries of Belgium, Czech…Official notice →
  6. US
    Drugs & medicalClass II

    Argon Medical Devices, Inc recalls Argon Medical Devices, REF 352506070E, Option Elite Retrievable Vena Cava Filter, Suitable of Jugular or Femoral Delivery, STERILEO,…

    There is the potential that the vena Cava Filter may not perform as intended which may result in pulmonary embolism.

    FDA · DeviceArgon Medical Devices, IncUS Nationwide distribution in the state of TX.Official notice →
  7. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls FRED Flow Re-Direction Endoluminal Device, REF: FRED3507-PMA, FRED3511, FRED3511-PMA, FRED3516-CA, FRED3516-PMA, FRED3524, FRED3524-CA,…

    Intracranial aneurysm flow diverter device may be manufactured with an incorrect inner stent length and may not perform as intended.

    FDA · DeviceMICROVENTION INC.Worldwide - US Nationwide distribution including in the states of AZ, NY, FL,…Official notice →
  8. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Ringed DxI Reaction Vessels (RVs)

    Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI,…Official notice →