Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 250-Intended for in vitro diagnostic use and is designed to…
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of…
Read the official notice →- Date
- 2023-12-27
- Country
- United States
- Brand
- Siemens Healthcare Diagnostics
- Product
- RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 250-Intended for in vitro diagnostic use and is designed to provide the…
- Classification
- Class II
- Units
- Nationwide with Worldwide Distribution to: Albania Algeria Angola Argentina…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Healthcare Diagnostics recalls
- 2024-03-27
US
Siemens Healthcare Diagnostics, Inc. recalls DCA Systems HbA1c Reagent Kit
The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the country of Taiwan.Official notice →- 2024-03-20
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Erythropoietin (EPO) Assay (100 Test)
Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of AL, CA, IA, IL, MI,…Official notice →- 2024-02-14
US
Siemens Healthcare Diagnostics, Inc. recalls Total immunoglobulin M (IgM total) IVD, kit, nephelometry/turbidimetry
Potential for a negative bias with quality control (QC) and patient sample results when using the Atellica CH Immunoglobulin M_2 (IgM_2) reagent.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →- 2023-12-27
US
Siemens Healthcare Diagnostics Inc recalls epoc NXS Host Blood Analysis System, Siemens Material Numbers 11413497 (EU), 11413475 (US), 11413506 (CA), 11413517 (MX),…
Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing erroneous results to be printed. When the results are run on the epoc NXS…
- 2023-12-06
US
Siemens Healthcare Diagnostics Inc recalls RAPIDPOINT 500 Blood Gas System
There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The…
FDA · DeviceSiemens Healthcare Diagnostics IncDomestic: Nationwide Distribution. Foreign: Albania, Algeria, Angola,…Official notice →- 2023-10-11
US
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Iron_2 (IRON_2) Reagents, Material Numbers 10377510 (6 x 350 tests) and 10341118 (7 x 145 tests)
There is a potential for a positive bias on Quality Control (QC) and patient sample results when using some IRON_2 reagent wedge pairs. When the issue occurs, the observed bias will vary wedge to wedge.
Other drugs & medical recalls in the United States
- 2024-01-03
US
CARDINAL HEALTHCARE recalls Enoxaparin Sodium Injection, USP, 150MG/ML Single-Dose Syringes with Automatic Safety Device, For Subcutaneous Injection, Ten 1 mL…
CGMP Deviations: Products were exposed to temperatures outside of the products labeled storage conditions.
US
GlaxoSmithKline LLC recalls ANORO ELLIPTA (umeclidinium and vilanterol inhalation powder) 62.5 mcg/25mcg. 1 inhaler contains 30 doses (60 blisters total), Rx Only,…
Failed Release Testing: Coarse Particle Mass for umeclidinium Out of Specification
- 2023-12-27
US
Stryker Spine recalls LITe Decompression Snake Arm, REF 48080230
Product arm was manufactured with the incorrect clamp component. As a result, the affected product arm cannot attach to the arm post shaft.
US
Maquet Medical Systems USA recalls Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i System, part numbers 701072780 and 701041802
Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be…
FDA · DeviceMaquet Medical Systems USADomestic distribution Nationwide. Foreign distribution to Argentina,…Official notice →US
Abbott Laboratories recalls Abbott CELL-DYN Ruby, Model CD-Ruby, List number 08H67-01, and Abbott CELL-DYN Sapphire, Model CD Sapphire, list number 08H00-01.…
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
FDA · DeviceAbbott LaboratoriesDistribution was made to AL, AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA,…Official notice →US
Zeus Scientific, Inc. recalls ZEUS ELISA Parvovirus B19 IgM Test System, Product Number SM9Z7701G
The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.