REPLIGEN CORPORATION recalls MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser…
Read the official notice →- Date
- 2026-01-21
- Country
- United States
- Brand
- REPLIGEN CORPORATION
- Product
- MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
- Classification
- Class II
- Units
- U.S.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2026-01-21
US
Boston Scientific Corporation recalls Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF…
Boston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation…
FDA · DeviceBoston Scientific CorporationUS Nationwide distribution in the states of MA, WV, OH, CA, MN, NY, VA, IL TX,…Official notice →US
Mobius Imaging, LLC recalls Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System.
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
FDA · DeviceMobius Imaging, LLCUS Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL,…Official notice →US
Medicrea International recalls Medicrea, IB3D PL Instruments Set, Rx Only REF: SPS03174
Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.
FDA · DeviceMedicrea InternationalUS Nationwide distribution in the states of TN, CA, MN, NC, TX, FL, OH, RI, VA.Official notice →US
Abbott Diabetes Care, Inc. recalls FreeStyle Libre 3 Sensor REF 72081-01 UDI-DI code: 00357599818005 The FreeStyle Libre 3 sensor is a component of the FreeStyle Libre 3…
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
FDA · DeviceAbbott Diabetes Care, Inc.U.S.: AK, AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO,…Official notice →US
Canon Medical System, USA, INC. recalls VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed…
FDA · DeviceCanon Medical System, USA, INC.U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD,…Official notice →US