QUIDEL CARDIOVASCULAR INC recalls Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01,…
Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is…
Read the official notice →- Date
- 2023-07-12
- Country
- United States
- Brand
- QUIDEL CARDIOVASCULAR INC
- Product
- Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01,…
- Classification
- Class I
- Units
- US: TX, NC, CA, LA, SC, OR, HI, OK, AZ, AL, KS, WA, FL, VA, IA, WV, MT, UT, CO…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2023-07-12
US
Intuitive Surgical, Inc. recalls da Vinci Xi/X Tip-Up Fenestrated Grasper, REF: 470347, VER: 12
Fenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or lower grips, which could lead to stainless steel fragments falling into the patient that may be either detected…
FDA · DeviceIntuitive Surgical, Inc.Worldwide - US Nationwide distribution in the states of PR, FL, MN, PA, AZ,…Official notice →US
Full Vision Inc recalls TRACKMASTER, TREADMILL GE T2100-ST2 220V CHINA, MODEL NUMBER 317-07927GE CHINA
Treadmill malfunction: drive PCB can misread the signals and stop or reverse and then accelerate forward.
FDA · DeviceFull Vision IncUSA (Nationwide), Australia, Bahrain, Canada, Chile, Colombia, Costa Rica,…Official notice →US
GE Healthcare Finland Oy recalls TruSignal Sensitive Skin Sensor, REF TS-SE-3; Oximeter
There is a potential reduction of energy reaching patient during defibrillation, potential contact with unintended voltage, or inaccurate measurement.
US
Strides Pharma Inc. recalls Losartan Potassium Tablets, USP, 25 mg, 1000 film coated tablets per bottle, Rx Only, Manufactured by: Vivimed Life Sciences Private…
Presence of Foreign Substance: Presence of a small piece of blue plastic embedded in the tablet.
US
US
Amerisource Health Services LLC recalls Tizanidine Tablets, USP, 4 mg, 100 Tablets (10 tablets x 10 unit dose blister packs) per carton, Rx Only, Packaged and…
Failed Dissolution Specifications: this repackaged product was recalled by the manufacturer, Dr. Reddy's Laboratories, Inc., due to out of specification results for dissolution.