Princeton Biomeditech Corp recalls Osom Ultra hCG Combo Test- IVD immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine for…
This recall was initiated due to customer complaints of false negative of hCG test results when running patient urine samples. Use of the above listed product lot may result in false negative results. An end user…
Read the official notice →- Date
- 2024-11-06
- Country
- United States
- Brand
- Princeton Biomeditech Corp
- Product
- Osom Ultra hCG Combo Test- IVD immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine for the…
- Classification
- Class II
- Units
- US Distribution to MO only.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2024-11-06
US
Integra LifeSciences Corp. recalls Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694
Out of specification endotoxin levels.
US
Ambu Inc. recalls Ambu¿ VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number…
Ambu has received complaints on Ambu¿ VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.
US
Trackx Technology Inc recalls Product Name: Insight Base Flex - 8, Ref 608-3TX; The device consists of a sterile Snap that includes reflective features that facilitate…
The patient label stickers may have the wrong expiration date recorded. The contract manufacturer entered the manufacturing date (2024-04-01) as the expiration date. The actual expiration date of the product is…
US
DRG International, Inc. recalls 17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone…
The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to…
FDA · DeviceDRG International, Inc.AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy…Official notice →US
Covidien LP recalls Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CFN CH06-40-75US), 7mm x 40mm…
Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.
FDA · DeviceCovidien LPWorldwide Distribution: US (Nationwide) and OUS (foreign) to countries of:…Official notice →US
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/2X2 Model/Catalog Number: 801406. The surgical patties and surgical strips are indicated for the use in protection of…
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
FDA · DeviceIntegra LifeSciences Corp.Worldwide Distribution. US Nationwide, Canada Germany United Kingdom of Great…Official notice →