Drugs & medicalClass II

Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc. recalls A1CNow+ Multi-Test HbA1c System consists of (1) a semi-disposable plastic-encased device (the…

The reason for this recall is PTS Diagnostics, Inc. received complaints of non-matching cartridge and analyzer codes within the product test system box. Use of cartridges and analyzers with non-matching codes may…

Read the official notice →
Date
2024-10-30
Country
United States
Brand
Polymer Technology Systems, Inc. dba PTS Diagnostics
Product
A1CNow+ Multi-Test HbA1c System consists of (1) a semi-disposable plastic-encased device (the monitor), (2) a plastic cartridge enclosing…
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Japan, South Africa.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Polymer Technology Systems, Inc. dba PTS Diagnostics recalls

  1. 2022-09-21
  2. US
    Drugs & medicalClass III

    Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc. recalls PTS Panels CHOL+GLU Test Strips (REF 1765)

    Incorrect instructions for use were packaged into the final product.

    FDA · DevicePolymer Technology Systems, Inc. dba PTS Diagnostics, Inc.Worldwide distribution - US Nationwide distribution in the state of Virginia…Official notice →
All 2 Polymer Technology Systems, Inc. dba PTS Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2024-10-30
  2. US
    Drugs & medicalClass II

    Boston Scientific Corporation recalls Boston Scientific myLUX Patient Application for an Apple Mobile device, Model Number 2939, Version 2.0.101, 2.0.110, cardiac…

    Patients using the Model 2939 myLUX Patient Application on their Apple mobile device (e.g., iPhone) may not be transmitting data from their Insertable Cardiac Monitor to the LATITUDE Clarity website.

    FDA · DeviceBoston Scientific CorporationNationwide distribution to AL, AZ, CA, CO, FL, GA, ID, KY, MA, MD, NC, NJ, NY,…Official notice →
  3. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate…

    Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon,…

    FDA · DeviceOlympus Corporation of the AmericasWorldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA,…Official notice →
  4. US
    Drugs & medicalClass II

    Spectrum Medical Ltd. recalls Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing.…

    The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.

    FDA · DeviceSpectrum Medical Ltd.US Nationwide distribution including in the states of Alabama, California,…Official notice →
  5. US
    Drugs & medicalClass II

    Ad-Tech Medical Instrument Corporation recalls AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Electrode

    Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol.…

    FDA · DeviceAd-Tech Medical Instrument CorporationUS Nationwide distribution in the states of CA, OH.Official notice →
  6. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Baxter Operating Table TruSystem 7000 (dV) V, Product code REF 1841083

    The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.

    FDA · DeviceBaxter Healthcare CorporationWorldwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Blue Ortho recalls Predict+ is a data-driven, clinical decision support tool that uses machine learning to provide predictions of individual patient outcomes…

    Predict+ is being recalled for lack of pre-market clearance/approval.

    FDA · DeviceBlue OrthoUS Nationwide distribution.Official notice →
See all drugs & medical recalls →