Drugs & medicalClass II

Philips North America recalls Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This…

Read the official notice →
Date
2023-11-22
Country
United States
Brand
Philips
Product
Spectral CT on Rails, Software Version 5.1.0, Model Number 728334
Classification
Class II
Units
US Nationwide distribution in the state of MN.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Philips recalls

  1. 2023-12-27
  2. US
    Drugs & medicalClass II

    Philips North America Llc recalls The device is a whole-body CT x-ray system. The acquired x-ray transmission data is reconstructed by computer into cross-sectional…

    Three software issues affecting incorrect image display, error interpreting patient images due to image reporting, and patient exposure to incorrect image/function during clinical use.

    FDA · DevicePhilips North America LlcDomestic: AL, AR, AZ, CA, FL, GA, HI, IL, IN, KY, LA, MD, ME, MI, MS, NC, NJ,…Official notice →
  3. 2023-12-20
  4. US
    Drugs & medicalClass I

    Philips North America Llc recalls Panorama 1.0T HFO, Magnetic Resonance (MR) System; Model Numbers 781250 and 781350. Philips Magnetic Resonance (MR) systems are Medical…

    Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain…

    FDA · DevicePhilips North America LlcDomestic distribution to AL AZ CA CO DE FL GA IA IL IN KS KY LA MA MD MI MO NE…Official notice →
  5. 2023-11-29
  6. US
    Drugs & medicalClass II

    Philips North America Llc recalls SmartPath to dStream for 1.5T

    Potential component failure in the Gradient Coil could product smoke and/or fire.

    FDA · DevicePhilips North America LlcDomestic: Nationwide Distribution; Foreign: Albania, Andorra, Argentina,…Official notice →
  7. 2023-11-22
  8. US
    Drugs & medicalClass II

    Philips North America recalls Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

    When preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be restarted.

    FDA · DevicePhilips North AmericaUS Nationwide distribution in the state of MN.Official notice →
  9. US
    Drugs & medicalClass II

    Philips North America Llc recalls Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for radiation oncology and…

    For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect…

    FDA · DevicePhilips North America LlcNationwide Foreign: Austria Belgium Canada China Estonia France Germany…Official notice →
  10. 2023-11-17
  11. SEEU
    Otherenvironment

    Remote control (Philips) · 3986GR08

    The solders in the product have an excessive concentration of lead (measured value up to 64 % by weight).Lead poses a risk to the environment. The product does not comply with the requirements of the Directive on the…

    EU Safety Gate · reported by SwedenPhilipsOfficial notice →
All 100 Philips recalls →

Other drugs & medical recalls in the United States

  1. 2023-11-29
  2. US
    Drugs & medicalClass III

    Glenmark Pharmaceuticals Inc., USA recalls Esomeprazole Magnesium Delayed-Release Capsules, USP 20mg, Packaged as (a) 30-count bottle, NDC 68462-390-30; (b) 1000-count…

    Failed Impurities/Degradation Specifications:Out of Specification result reported for the test of organic impurities for the drug product,at the 18 month time point in long term stability study (25¿C/60% RH).

    FDA · DrugGlenmark Pharmaceuticals Inc., USANationwide in the USAOfficial notice →
  3. US
    Drugs & medicalClass II

    Integrity Bio-Chemicals LLC recalls Integrity Biochem HSC 70-VA, Alcohol Antiseptic 70%, Hand Sanitizer, Vanilla Scented Topical Gel 208.19L (55 gallons)- drum.…

    CGMP Deviation: Third party test results showing a presence of acetal and acetaldehyde at levels above USP specified amounts.

    FDA · DrugIntegrity Bio-Chemicals LLCDistributed to one distributor in OK with possibility of further distributionOfficial notice →
  4. US
    Drugs & medicalClass I

    The Harvard Drug Group LLC dba Major Pharmaceuticals and… recalls Polyvinyl Alcohol 1.4% Lubricating Eye Drops, packaged in 0.5 FL OZ (15mL) bottles, Dist. by: RUGBY…

    Non-Sterility: FDA found insanitary conditions and positive bacterial test results from environmental sampling at the manufacturing facility.

    FDA · DrugThe Harvard Drug Group LLC dba Major Pharmaceuticals and…Nationwide within the United StatesOfficial notice →
  5. 2023-11-22
  6. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Express Dry Seal Chest Drains (Express): 4000-100N DRAIN, EXPRESS, SINGLE; 4050-100N DRAIN, EXPRESS, BRU

    Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection…

    FDA · DeviceAtrium Medical CorporationWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →
  7. US
    Drugs & medicalClass II

    ROi CPS LLC recalls Regard CV PK, Item Numbers a) 880289014, b) 880289015; cardiovascular convenience kit

    Specific Regard kits contain the Gentige Chest Drain Products being recalled by Gentige. The Instructions for Use (IFU) for the chest drains do not provide sufficient precaution instruction for proper set up of…

    FDA · DeviceROi CPS LLCMOOfficial notice →
  8. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls ADLT BLOOD CULTURE COLLECT KIT, REF DYNDH1245

    This recall is being issued due to items within the kit have expired prior to the overall kit expiration date listed on the insert.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the state of California.Official notice →