PHILIPS MEDICAL SYSTEMS recalls Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333…
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
Read the official notice →- Date
- 2026-10-14
- Country
- United States
- Brand
- Philips Medical Systems
- Product
- Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333…
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CO, FL,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips Medical Systems recalls
- 2026-08-05
US
PHILIPS MEDICAL SYSTEMS recalls IQon Spectral CT IQon Spectral CT Upgrades
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being…
FDA · DevicePHILIPS MEDICAL SYSTEMSUS Nationwide Distribution including Puerto RicoOfficial notice →- 2026-04-22
US
PHILIPS MEDICAL SYSTEMS recalls Philips Spectral CT on Rails. Model Number: 728334.
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of…
FDA · DevicePHILIPS MEDICAL SYSTEMSWorldwide - US Nationwide distribution in the states of Minnesota and the…Official notice →- 2025-11-05
US
PHILIPS MEDICAL SYSTEMS recalls Model: Incisive CT for Brazil SKD; Product Code (REF): 728146; Software Versions: 5.1.0.X & 5.1.1.X;
Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/…
FDA · DevicePHILIPS MEDICAL SYSTEMSDomestic: AL, AR, CA, CT, FL, GA, IL, IN, KY, MD, MI, OH, PA, PR, TN, TX, UT,…Official notice →- 2025-09-24
US
Philips Medical Systems (Cleveland) Inc recalls Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code:…
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
FDA · DevicePhilips Medical Systems (Cleveland) IncU.S Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT,…Official notice →- 2025-08-13
US
Philips Medical Systems (Cleveland) Inc recalls Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
FDA · DevicePhilips Medical Systems (Cleveland) IncWorldwide - US Nationwide distribution in the state of WV and the countries of…Official notice →- 2024-02-21
US
PHILIPS MEDICAL SYSTEMS recalls CombiDiagnost R90 1.1 (709031)
While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image…
FDA · DevicePHILIPS MEDICAL SYSTEMSUS Nationwide distribution in the states of CA, FL, GA, IL, IN, KS, MD, ME,…Official notice →
Other drugs & medical recalls in the United States
- 2026-10-14
US
Safeguard US Operating LLC recalls Battle Bandage, REF: 32-700; Patrol IFAK Bag, Black - Advanced, REF: 81-125-BL; Patrol IFAK Bag, Multicam - Advanced, REF: 81-125-MC;…
Compression bandage may be degraded, which may cause unwanted foreign material to be introduced inside the wound, which may lead to wound exploration and irrigation, formation of scar tissue, and infection.
FDA · DeviceSafeguard US Operating LLCWorldwide - US Nationwide distribution including in the states of FL, OK, CA,…Official notice →US
Limitless Cryo Spa LLC recalls Glypto Device, Glypto Cryo Machine;
Lack of 510(k) clearance.
US
Cardinal Health 200, LLC recalls Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J…
Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a…
US
GRI Medical & Electronic Technology Co., Ltd. recalls HALYARD Laparoscopy Comprehensive Pack II REF: 49332
Laparoscopy packs labeled as "latex free" were found to contain a latex urethral catheter
FDA · DeviceGRI Medical & Electronic Technology Co., Ltd.US Nationwide distribution in the state of MS.Official notice →US
QUANTUM OPS Inc recalls Brand Name: ADAPTABLE Beach Chair Product Name: Beach Chair surgical patient positioner Model/Catalog Number: 804-33-300 (Enovis);…
Due to a manufacturing issue (inadequate weld) for patient head positioning system.
FDA · DeviceQUANTUM OPS IncU.S. Nationwide distribution in the states of CA and TX.Official notice →US
Steris Corporation recalls VERIFY Biological Steam Test Pack Model/Catalog Number: S3065
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological…