Drugs & medicalClass II

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Zenition 50, Model #718096

Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used…

Read the official notice →
Date
2022-02-23
Country
United States
Brand
Philips Medical Systems Nederland
Product
Zenition 50, Model #718096
Classification
Class II
Units
Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Philips Medical Systems Nederland recalls

  1. 2023-09-13
  2. US
    Drugs & medicalClass II

    Philips Medical Systems Nederland B.V. recalls Allura Xper series

    A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation

    FDA · DevicePhilips Medical Systems Nederland B.V.US Nationwide DistributionOfficial notice →
  3. 2023-09-06
  4. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Wireless Foot Switch distributed with Philips Allura Xper and Azurion Interventional Fluoroscopic X-Ray Systems

    Loss of availability of the wireless foot switch during procedures.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  5. 2023-08-09
  6. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20. X-ray Image guided Therapy System.

    The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  7. 2023-06-07
  8. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1)…

    Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Nationwide Foreign: Algeria Argentina Armenia Australia Austria Bahrain…Official notice →
  9. 2023-04-26
  10. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls SmartPerfusion, release R1.0, R1.1, R1.1.1, R1.1.5, R1.1.5.1 and R1.1.6

    There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution.Official notice →
  11. 2022-05-11
  12. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion systems with software release R1.x

    In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.US Nationwide and Worldwide DistributionOfficial notice →
All 59 Philips Medical Systems Nederland recalls →

Other drugs & medical recalls in the United States

  1. 2022-02-23
  2. US
    Drugs & medicalClass II

    Smiths Medical ASD Inc. recalls Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time…

    There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the…

    FDA · DeviceSmiths Medical ASD Inc.Distribution US nationwide, including PR, VI, GU, and MP. There was…Official notice →
  3. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls Exeter V40 Cemented Hip (150mm) Stem, Part Number 0580-1-440

    There is a potential label mix between certain lots of Exeter V40 Cemented Hip (125mm) Stem and Exeter V40 Cemented Hip (150mm) Stem.

    FDA · DeviceHowmedica Osteonics Corp.The product was distributed to the following US states: MIOfficial notice →
  4. US
    Drugs & medicalClass II

    Merete Medical GmbH recalls Merete PediatrOS RigidTack, REF CP20120, Size 20 mm, Sterile, Rx Only

    Incorrect marketing label (25 mm) was applied to product size 20 mm.

    FDA · DeviceMerete Medical GmbHDomestic: Illinois; Foreign: Germany.Official notice →
  5. US
    Drugs & medicalClass II

    Oscor Inc. recalls Destino Twist 14F (also branded as Guidestar 14F)

    There is potential for separation between the threaded hub and threaded cap at the proximal end of the device.

    FDA · DeviceOscor Inc.Distribution to TN and PA in USA OUS distribution to Czech RepublicOfficial notice →
  6. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls 32MM -4 V40 TAPER VIT HEAD, Catalog Number 6260-5-032

    There is a potential for the outer Tyvek lid to de-bond from the sealed package.

    FDA · DeviceHowmedica Osteonics Corp.Worldwide distribution - US Puerto Rico and the countries of Canada, and…Official notice →
  7. US
    Drugs & medicalClass II

    Envisiontec US Llc recalls FLEXCERA SMILE -a Dental resins for the fabrication of artificial teeth

    Manufactured in a non-FDA-registered manufacturing facility and product mislabeled as being of German origin

    FDA · DeviceEnvisiontec US LlcUS Nationwide distribution.Official notice →
See all drugs & medical recalls →