Onkos Surgical, Inc. recalls a. Segmental Stem, Cemented, Straight, 100MM, 1 EACH; Model Numbers: CS-09100-03M, CS-10100-03M. b. Segmental Stem, Cemented, Straight,…
Potential for breach of Tyvek seals.
Read the official notice →- Date
- 2025-06-04
- Country
- United States
- Brand
- Onkos Surgical
- Product
- a. Segmental Stem, Cemented, Straight, 100MM, 1 EACH; Model Numbers: CS-09100-03M, CS-10100-03M. b. Segmental Stem, Cemented, Straight,…
- Classification
- Class II
- Units
- Product was shipped from Onkos' distribution in Olive Branch, MS to consignees…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Onkos Surgical recalls
- 2026-10-14
US
Onkos Surgical, Inc. recalls Onkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M
Personalized pelvic implants inadvertently mixed up and distributed to incorrect customers.
FDA · DeviceOnkos Surgical, Inc.US Nationwide distribution in the states of North Carolina, and California.Official notice →- 2025-09-03
US
Onkos Surgical, Inc. recalls ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS STEM EXTENSION, STRAIGHT, SPLINED, SLOTTED. Canal filling stem implants.
Unidentified substance of unknown impact was identified on devices during a retrospective evaluation.
FDA · DeviceOnkos Surgical, Inc.US Nationwide distribution in the states of AK, AZ, CA, FL, GA, IL, IN, KS,…Official notice →- 2025-07-09
US
Onkos Surgical, Inc. recalls My3D Personalized Solutions Humeral Cup, Model Number: C24-0223-0003. It is a component of a Custom Constrained Shoulder Arthroplasty Device
Required inspections were not performed on finished product prior to release and distribution.
FDA · DeviceOnkos Surgical, Inc.US Nationwide distribution in the state of Maryland.Official notice →- 2025-06-04
US
Onkos Surgical, Inc. recalls ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System
Potential challenge with intraoperative assembly of the external taper of a single midsection with the internal taper of a mating proximal femur.
FDA · DeviceOnkos Surgical, Inc.US Nationwide distribution in the states of AZ, CA, FL, FL, MD, MO, NC, TN.Official notice →US
Onkos Surgical, Inc. recalls ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with BioGrip.
Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems…
FDA · DeviceOnkos Surgical, Inc.US Nationwide distribution in the states of California, Kentucky, and Wisconsin.Official notice →
Other drugs & medical recalls in the United States
- 2025-06-11
US
Apotex Corp. recalls Lacosamide Oral Solution, USP, 10 mg/mL, 200 mL, Rx Only, Manufactured by: Apotex Inc., Toronto, Ontario, Canada M9L 1T9, Manufactured…
Defective Container: This recall is being initiated due to a leaking unit stored horizontally.
US
Torrent Pharma Inc. recalls Fluoxetine Tablets, USP, 20 mg, 28-count Carton (4x7 Unit-dose), Manufactured by: TORRENT PHARMACEUTIALS LTD., Indrad-382 721 INDIA;…
CGMP Deviations: Presence of N-Nitroso Fluoxetine exceeding interim acceptable intake limit.
US
- 2025-06-04
US
BioFire Diagnostics, LLC recalls BioFire Respiratory Panel 2.1 plus Kit, REF: 423740, For Filmarray systems
Due to manufacturing error, respiratory panel may result in elevated control failures and/or false negative results.
FDA · DeviceBioFire Diagnostics, LLCInternational distribution in the country of Singapore.Official notice →US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides.
Product was incorrectly assigned GEN 88 instead of GEN 83 on product carton, foil wrapper, and cartridge (cart) label.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of AR, AZ, CO, FL, GA,…Official notice →US
Cardinal Health 200, LLC recalls Catalog No. 8888160119; Umbilical Vessel Catheter Insertion Tray with 2.5 and 5 French UVCs
Packaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates.
FDA · DeviceCardinal Health 200, LLCWorldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →