NOXBOX LTD recalls NOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I.
Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.
Read the official notice →- Date
- 2026-10-07
- Country
- United States
- Brand
- Noxbox
- Product
- NOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I.
- Classification
- Class II
- Units
- US Nationwide distribution in the state of Georgia.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Noxbox recalls
- 2026-09-23
US
NOXBOX LTD recalls NOxBOXi Nitric Oxide Delivery System.
Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.
- 2025-10-15
US
NOXBOX LTD recalls NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
US
NOXBOX LTD recalls NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have…
- 2025-09-17
US
NOXBOX LTD recalls NOxBOXi Nitric Oxide Delivery System
Replacement of the internal Sample Pump due to high failure rates.
- 2023-06-21
US
NOXBOX LTD recalls NOxBOXi Nitric Oxide Delivery System, NOXBOX-I, REQNOXBOX-I, Nitric Oxide Delivery Device
Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or…
FDA · DeviceNOXBOX LTDWorldwide distribution - US Nationwide and the countries of Bahrain, Bolivia,…Official notice →
Other drugs & medical recalls in the United States
- 2026-10-07
US
Micro Therapeutics, Inc. recalls Pipeline Flex Embolization Device with Shield Technology REF PED2-300-12 UDI-DI code: 00763000448806 he Pipeline Flex Embolization…
Due to incorrect label/product dimensions. Product is supposed to be 3.00 mm in diameter, but unit is packaging was a 2.50 mm in diameter.
FDA · DeviceMicro Therapeutics, Inc.U.S. Nationwide distribution in the states of GA and TX.Official notice →US
Percussionaire Corporation recalls Intrapulmonary Percussive Ventilation, IPV HC, Product Number: S00009; Phasitron, P5-HC
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents,…
FDA · DevicePercussionaire CorporationWorldwide - US Nationwide distribution including in the states of CA, PA, OK,…Official notice →US
Philips North America recalls Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software…
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction…
FDA · DevicePhilips North AmericaWorldwide distribution - Nationwide and the countries of Algeria, Argentina,…Official notice →US
Merit Medical Systems, Inc. recalls Embosphere 500-700 micrometer 2 mL Syringe Sterile Steam, REF: S420GH/EUA, S620GH/EUA; Embosphere 700-900 micrometer 2 mL Syringe…
Microspheres used in embolization of tumors, prostatic arteries, arteriovenous malformations, and blood vessels may have holes in packaging that may affect the sterile barrier, which if used could lead to infection.
FDA · DeviceMerit Medical Systems, Inc.Worldwide - US Nationwide distribution in the states of CA, GA and the…Official notice →US
Heraeus Medical GmbH (Dental Division) recalls PALACOS MV+G pro 40; PALACOS MV+G pro 80. Model/Catalog Number: 5081290; 5081291. Radiopaque, poly(methyl…
Potential for monomer liquid leakage.
FDA · DeviceHeraeus Medical GmbH (Dental Division)Worldwide distribution - US Nationwide and the countries of Aland Islands,…Official notice →US
Olympus Corporation of the Americas recalls Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
FDA · DeviceOlympus Corporation of the AmericasInternational distribution to the country of Japan.Official notice →