Drugs & medicalClass II

Mirion Technologies (Capintec), Inc. recalls Captus 4000e Thyroid Uptake System. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests.

Complaints of unexpected detachment of the collimator have been reported.

Read the official notice →
Date
2024-06-05
Country
United States
Brand
Mirion Technologies
Product
Captus 4000e Thyroid Uptake System. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests.
Classification
Class II
Units
Worldwide - US Nationwide distribution and the countries of ALGERIA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Mirion Technologies recalls

  1. 2026-09-23
  2. US
    Drugs & medicalClass II

    Mirion Technologies (Capintec), Inc. recalls Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

    Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.

    FDA · DeviceMirion Technologies (Capintec), Inc.Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA,…Official notice →
  3. 2024-02-21
  4. US
    Drugs & medicalClass II

    Mirion Technologies (Capintec), Inc. recalls Captus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or Nuclear Medicine…

    Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel, causing potential injury with a patient or operator

    FDA · DeviceMirion Technologies (Capintec), Inc.Nationwide Foreign: AUSTRIA BANGLADESH BRASIL Brazil CANADA CHILE Costa Rica…Official notice →
All 3 Mirion Technologies recalls →

Other drugs & medical recalls in the United States

  1. 2024-06-05
  2. US
    Drugs & medicalClass II

    Lucid Diagnostics, Inc. recalls Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter, 10 pores/inch) Model/Catalog Number: EC25T10VI Software Version: N/A…

    The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper…

    FDA · DeviceLucid Diagnostics, Inc.US Nationwide distribution in the states of MN, MD.Official notice →
  3. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column,…

    Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

    FDA · DeviceFUJIFILM Healthcare Americas CorporationUS Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.Official notice →
  4. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper FD10 system with an ADN7NT patient table. Model Number: 722010, Software Version Number 8.1.100.

    When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Domestic: OK International: India, Spain, U.A.E.Official notice →
  5. US
    Drugs & medicalClass II

    Covidien, LP recalls Tri-Staple 2.0 Black Reinforced Intelligent Reload- Applications in abdominal, gynecologic, pediatric and thoracic surgery for resection,…

    Potential damage to the cartridge can result in poor staple formation and/or incomplete staple line, may cause a delay in surgery

    FDA · DeviceCovidien, LPWorldwide - US Nationwide distribution in the state of NJ and the countries of…Official notice →
  6. US
    Drugs & medicalClass II

    Boston Scientific Neuromodulation Corporation recalls Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32…

    Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients…

    FDA · DeviceBoston Scientific Neuromodulation CorporationWorldwide - US Nationwide distribution including in the states of TX, WA, MT,…Official notice →
  7. US
    Drugs & medicalClass I

    MEDLINE INDUSTRIES, LP - Northfield recalls Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm ID; (2) REF…

    Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were also reports the…

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldDistribution was nationwide. There was government/military distributon.…Official notice →
See all drugs & medical recalls →