Drugs & medicalClass II

Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896

A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53…

Read the official notice →
Date
2026-07-15
Country
United States
Brand
Medtronic MiniMed
Product
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
Classification
Class II
Units
International distribution to the countries of Spain, Great Britain,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Medtronic MiniMed recalls

  1. 2026-10-14
  2. US
    Drugs & medicalClass II

    Medtronic MiniMed, Inc. recalls Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The…

    Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The…

    FDA · DeviceMedtronic MiniMed, Inc.U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI,…Official notice →
  3. 2026-08-26
  4. US
    Drugs & medicalClass II

    Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW

    Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing…

    FDA · DeviceMedtronic MiniMed, Inc.International distribution to the countries of Germany, Denmark, Austria,…Official notice →
  5. 2026-07-29
  6. US
    Drugs & medicalClass II

    Medtronic MiniMed, Inc. recalls MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)

    For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This…

    FDA · DeviceMedtronic MiniMed, Inc.Worldwide - US Nationwide distribution in the states and territories of PR,…Official notice →
  7. 2026-04-08
  8. US
    Drugs & medicalClass II

    Medtronic MiniMed, Inc. recalls MiniMed 720G Insulin Pump (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)

    All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the…

    FDA · DeviceMedtronic MiniMed, Inc.Worldwide - US Nationwide distribution including in the states of OR, CA, AZ,…Official notice →
  9. 2026-04-01
  10. US
    Drugs & medicalClass II

    Medtronic MiniMed, Inc. recalls MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61

    Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or…

    FDA · DeviceMedtronic MiniMed, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
  11. 2026-02-11
  12. US
    Drugs & medicalClass II

    Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for…

    A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.

    FDA · DeviceMedtronic MiniMed, Inc.Worldwide - U.S. National distribution including in the states of AL, AR, AZ,…Official notice →
All 19 Medtronic MiniMed recalls →

Other drugs & medical recalls in the United States

  1. 2026-07-15
  2. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls 5008X Blood Tubing Set, 24 units per case. Includes the below model numbers and descriptions. 1. 5008X HD Pre-Flush…

    Potential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the 5008X CAREsystem.

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, GA,…Official notice →
  3. US
    Drugs & medicalClass II

    Encore Medical, LP recalls Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 342-19-703 Product Description: Enovis…

    Due to packaging discrepancy with incorrect product size mixed up.

    FDA · DeviceEncore Medical, LPGL112 Certus Medical LLC Indiana (IN) & Michigan (MI) 342-19-703 177T1057…Official notice →
  4. US
    Drugs & medicalClass II

    Klowen Braces recalls KL Owen Custom Orthodontic Solution, model/catalog number CIAS-800033,

    A product error defect may cause incorrect bracket placement.

    FDA · DeviceKlowen BracesUS Nationwide distribution in the states of OH, TX, SC VA.Official notice →
  5. US
    Drugs & medicalClass II

    Medline Industries, LP recalls Medline Convenience Kits: 1) ANESTHESIA KIT PEDS 90-LF, Model Number: DYNJAP10125A

    Convenience Kits contain Covidien Mon-a-Therm General Purpose Temperature Probes Size 9 Fr. Medtronic has issued a recall of specific production lots of the Covidien Mon-a-Therm general purpose temperature probe-size 9…

    FDA · DeviceMedline Industries, LPUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Boston Scientific Neuromodulation Corporation recalls Intracept Access Instruments, Description/REF: INTRACEPT ACCESS INSTRUMENTS- DRILL/RLV0035, INTRACEPT ACCESS…

    Intraosseous nerve ablation system instruments may have sterile barrier breach of the outer tray seal, which may introduce a risk of sterility breach, and if breached could lead to severe infection.

    FDA · DeviceBoston Scientific Neuromodulation CorporationWorldwide - US Nationwide distribution including in the states of NJ, WI, MI,…Official notice →
  7. US
    Drugs & medicalClass II

    ConvaTec, Inc recalls Aquacel Ag Surgical SP REF: 413556 9cm x 15cm/3.5 in. x 6 in., Sterile R

    Surgical dressing manufactured out of specification (greater thickness).

    FDA · DeviceConvaTec, IncUS Nationwide distribution in the states of MD, OH, IL, NC, AZ, NY, MN, TX,…Official notice →