Drugs & medicalClass II

MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE Polyurethane Foam Bedside Cleaning Kit, REF DYK500FS

Sporadic reports were received of discoloration on the foam sponge on certain lots of the product, which has been identified as microbial contamination.

Read the official notice →
Date
2022-06-08
Country
United States
Brand
MEDLINE INDUSTRIES, LP - Northfield
Product
MEDLINE Polyurethane Foam Bedside Cleaning Kit, REF DYK500FS
Classification
Class II
Units
US and Panama
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More MEDLINE INDUSTRIES, LP - Northfield recalls

  1. 2022-10-05
  2. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner…

    Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the…

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldNationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID,…Official notice →
  3. 2022-08-10
  4. US
  5. 2022-07-06
  6. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Medline Pill Splitters; 1 each/box; 144 boxes/carton

    Product lid (which houses a blade) is coming unhinged from the base, thus exposing the blade.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Singapore,…Official notice →
  7. 2022-06-01
  8. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls ABX Pentra Calcium AS CP, Reagent 90 mL, REF A11A01954

    Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this impacted Calcium reagent will not be provided due to…

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Distribution to North Carolina.Official notice →
  9. 2022-04-20
  10. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Medline Standard PVC Laryngeal Masks, 1 device per pouch, 5 each pouches per box

    The mask cuff may disconnect from the device's breathing tube.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution.Official notice →
  11. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Stanbio TDM/Beta-Hydroxybutyrate Linearity Standards, packaged in boxes containing Levels 1-6, 1 x 4 mL each, REF 2450-604,…

    The product was stored incorrectly due to improper storage controls which may lead to delayed results.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldDistribution was made to AL, FL, NE OH, TX, WA, and WI. There was no…Official notice →
All 117 MEDLINE INDUSTRIES, LP - Northfield recalls →

Other drugs & medical recalls in the United States

  1. 2022-06-08
  2. US
    Drugs & medicalClass II

    MicroPort Orthopedics Inc. recalls PROFEMUR PLASMA Z CLASSIC STEM SIZE 7 STANDARD SHORT NECK

    One manufacturing lot is missing the plasma-spray coating in the proximal region of the stem body.

    FDA · DeviceMicroPort Orthopedics Inc.Worldwide distribution - US Nationwide distribution in the states of FL, IN,…Official notice →
  3. US
    Drugs & medicalClass II

    Medtronic Xomed, Inc. recalls NIM Vital Patient Interface 4.0 (P/N: NIM4CPB1) NIM with Software version 1.3.2

    Software anomaly was identified.

    FDA · DeviceMedtronic Xomed, Inc.Worldwide distribution - US Nationwide distribution in the states of TX, CO,…Official notice →
  4. US
    Drugs & medicalClass II

    Dynex Technologies, Inc. recalls DYNEX Agility, Agility Analyzer, Model No. 67000

    Control samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's SmartKit will be used instead of the…

    FDA · DeviceDynex Technologies, Inc.Worldwide distribution - US Nationwide distribution in the states of AL, CA,…Official notice →
  5. US
    Drugs & medicalClass II

    Tornier, Inc recalls Ratcheting Screwdriver Handle, provided in the Aequalis Perform Reversed kit

    The device is unable to meet sterilization requirements when sterilized per instructions in an Aesculap container.

    FDA · DeviceTornier, IncUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Becton, Dickinson and Company, BD Biosciences recalls BD Anti-Kappa PE (Cat. No. 346587) is an analyte specific reagent for the Anti-Kappa antibody, which specifically…

    Anti-Kappa APC & Anti-Kappa PE analyte reagents may exhibit uncharacteristic flow profiles.

    FDA · DeviceBecton, Dickinson and Company, BD BiosciencesWorldwide distribution - US Nationwide distribution in the states of AL, CA,…Official notice →
  7. US
    Drugs & medicalClass II

    Abbott recalls FIRMap" Catheter, 60mm Basket

    Incorrect product labeling.

    FDA · DeviceAbbottUS Nationwide Distribution to states of: MO, CA, OH, WA and OUS…Official notice →