Mckesson Medical-Surgical Inc. Corporate Office recalls SEKISUI DIAGNOSTICS LLC SERUM HCG OSOMGENZME CONTROL SET.
transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an…
Read the official notice →- Date
- 2025-06-04
- Country
- United States
- Brand
- Mckesson Medical-Surgical Inc. Corporate Office
- Product
- SEKISUI DIAGNOSTICS LLC SERUM HCG OSOMGENZME CONTROL SET.
- Classification
- Class II
- Units
- US Nationwide distribution in the states of AL, AR, IL, KS, MT, NC, NJ, OR,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Mckesson Medical-Surgical Inc. Corporate Office recalls
- 2025-08-27
US
Mckesson Medical-Surgical Inc. Corporate Office recalls Clearview Strep A Exact II Dipstick Respiratory Kit, CLIA Waived. Catalog Number: 4581225020
Due to a classification error in the inventory management system, specific medical devices were shipped to customers that did not have the requisite license to purchase such devices.
FDA · DeviceMckesson Medical-Surgical Inc. Corporate OfficeUS Nationwide distribution in the state of Georgia.Official notice →- 2025-07-30
US
Mckesson Medical-Surgical Inc. Corporate Office recalls EVENITY, (romosozumab -aqqg) injection, 105mg/1.17 mL, For Subcutaneous Use Only, Rx Only, Manufactured by:…
CGMP Deviations; potential temperature excursions due to transit delays
FDA · DrugMckesson Medical-Surgical Inc. Corporate OfficeWithin the U.S - OH, VA, FL.Official notice →- 2025-05-07
US
Mckesson Medical-Surgical Inc. Corporate Office recalls Infuvite Adult, multiple vitamins injection, Baxter, Five of Vial 1 (5 mL) and Five of Vial 2 (5mL), Rx Only,…
Severe thunderstorms caused transit delays of certain cold chain products that were shipped on 02-Apr-2025 and delivered after the 48-hours specified delivery time, on 07-Apr-2025 and 08-Apr-2025. The affected products…
FDA · DrugMckesson Medical-Surgical Inc. Corporate OfficeDistributed to Medical Facilities in MS and FL.Official notice →- 2024-12-18
US
- 2024-10-30
US
Mckesson Medical-Surgical Inc. Corporate Office recalls Quantimetrix Urine Chemistry Control Set Dipper Urinalysis Dipstick Testing Levels 1 & 2 6 X 15 mL-IVD intended…
The Quantimetrix Dipper Urinalysis Dipstick control was shipped without the validated summer packaging and may have been exposed to temperatures above 25¿C. This may lead to delay in patients' results.
FDA · DeviceMckesson Medical-Surgical Inc. Corporate OfficeUS Nationwide distribution in the state of DC.Official notice →- 2024-10-02
US
Mckesson Medical-Surgical Inc. Corporate Office recalls Proparacaine Hydrochloride Ophthalmic Solution, USP 0.5% (sterile), 15 mL bottle, Rx only, MFG: Bausch & Lomb,…
Temperature abuse: Products exposed to improper temperature above drug label specifications.
FDA · DrugMckesson Medical-Surgical Inc. Corporate OfficeProduct was distributed to two medical facilities in MD and VA.Official notice →
Other drugs & medical recalls in the United States
- 2025-06-04
US
C-RAD POSITIONING AB recalls Brand Name: Catalyst+ Product Name: C4D software used in conjunction with Catalyst (SP001-0025), Catalyst HD (SP001-0026), Catalyst PT…
Software issue with scanning equipment that can results in the filed rotation not being applied correctly.
FDA · DeviceC-RAD POSITIONING ABWorldwide - US Nationwide distribution in the states of AK, FL, WA, VA, WA,…Official notice →US
Onkos Surgical, Inc. recalls ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System
Potential challenge with intraoperative assembly of the external taper of a single midsection with the internal taper of a mating proximal femur.
FDA · DeviceOnkos Surgical, Inc.US Nationwide distribution in the states of AZ, CA, FL, FL, MD, MO, NC, TN.Official notice →US
Prismatik Dentalcraft, Inc. recalls Glidewell HT Implant Twist Drill ¿1.5 x 8 mm " SKU # 70-1071-SRG0266 UDI code: +D7457071SRG02660/$$76254798/16D20250219V The…
Due to packaging mix up. Package of finished devices were reported to have an incorrect finished device package inside.
FDA · DevicePrismatik Dentalcraft, Inc.U.S Nationwide distribution in the states of KY, MA, MI, NC, VA, and WV.Official notice →US
Maquet Cardiopulmonary Ag recalls Cardiohelp System HKH 8820 Wall Holder. Mounting bracket designed to hold the CARDIOHELP-i system for interhospital transport by…
HKH 8820 Wall Holder does not comply with standard DIN EN 1789:20 and has the potential to detach.
FDA · DeviceMaquet Cardiopulmonary AgDomestic: IL & MO. Foreign: Australia, Austria, Belgium, Canada, China,…Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →US
Onkos Surgical, Inc. recalls ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with BioGrip.