Luminex Corporation recalls VERIGENE Nucleic Acid Test Kit-BC-GN Extraction Tray - 9.73 mL, REF 20-009-021
Higher than normal "No Call Internal Control failures" due to a missing reagent within a small subset of the extraction tray lot.
Read the official notice →- Date
- 2022-05-25
- Country
- United States
- Brand
- Luminex
- Product
- VERIGENE Nucleic Acid Test Kit-BC-GN Extraction Tray - 9.73 mL, REF 20-009-021
- Classification
- Class II
- Units
- US
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Luminex recalls
- 2022-11-02
US
Luminex Corporation recalls ARIES SARS-CoV-2 Assay, REF: 50-10047
A part defect with the wash 1 syringe of an SARS-CoV-2 assay could lead to incomplete extraction thereby producing an invalid or false negative assay result.
FDA · DeviceLuminex CorporationDistribution US nationwide, Indonesia, and Hong Kong.Official notice →- 2022-08-24
US
Luminex Corporation recalls ARIES SARS-CoV-2 Assay
SARS-CoV-2 assay that detects SARS-CoV-2 ORF1ab and N genes may not detect the ORF1ab gene. A positive results if either SARS-CoV-2 genes is detected, but If both genes are mistakenly not detected, a false negative…
FDA · DeviceLuminex CorporationU.S. Nationwide distribution in the states of NE, TX, MS, TN, KY, ME, CA.Official notice →- 2022-07-13
US
Luminex Corporation recalls VERIGENE Enteric Pathogens Nucleic Acid Test, Part No. 30-002-23
There is a potential for false negative results using VERIGENE CDF Stool PREP KIT and VERIGENE EP Stool PREP KIT due to hydrophobic characteristics of the swab.
FDA · DeviceLuminex CorporationWorldwide distribution - US Nationwide and the countries of Kuwait, Austria,…Official notice →- 2022-05-25
US
Luminex Corporation recalls Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system
Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.
FDA · DeviceLuminex CorporationUS:OH, ME, PA, FL, IL, NE, IN, TX, GA, WA, MI, CO, AZ, OH, KY, CA, VT OUS: NoneOfficial notice →- 2022-04-20
US
Luminex Corporation recalls Verigene Processor SP, Model Numbers: 10-0000-07 and 10-0000-07-R
Two of three complaints were initiated due to the customers repeatedly obtaining No Call results on the Processor SPs in question.
- 2022-02-09
US
Luminex Corporation recalls ARIES HSV 1&2 Assay, Catalog/REF# 50-10017
Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.
FDA · DeviceLuminex CorporationWorldwide distribution - US Nationwide distribution in the states of New…Official notice →
Other drugs & medical recalls in the United States
- 2022-05-25
US
Ortho-Clinical Diagnostics, Inc recalls VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198
Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at…
FDA · DeviceOrtho-Clinical Diagnostics, IncWorldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda…Official notice →US
TURBETT SURGICAL, LLC recalls Turbett Surgical Container, TS1500
Eight units manufactured before June 2019 have exhibited weld failures breaching the sterile barrier after prolonged use.
US
Medtronic Inc recalls Instructions for Use, Patient Manuals, and Emergency Responder Guides for HeartWare Ventricular Assist Device (HVAD) System. HVAD System.…
Errors and inconsistencies that were identified in the Instruction for Use (IFU), Emergency Responder Guide (ERG), and Patient Manual (PM) for the HeartWare Ventricular Assist Device (HVAD) system.
FDA · DeviceMedtronic IncWorldwide Distribution: US (nationwide) and OUS to countries of: Australia,…Official notice →US
CooperSurgical, Inc. recalls VITRIGUARD, CARRIER, CLEAR; Model. No. 43782001 The VitriGuard device is a cryopreservation storage device composed of a stick and cap.
Products in this lot may have compromised barrier seals, introducing a risk to sterility.
FDA · DeviceCooperSurgical, Inc.Domestic distribution to AZ, CA, FL, IL, MI, MO, NY, OR, TX, and WV. No…Official notice →US
Heartware, Inc. recalls HeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1407DE Kit
Finnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and Emergency Responder Guide for the HeartWare Ventricular Assist Device (HVAD) system.
US
Windstone Medical Packaging, Inc. recalls Prevantics swabs and/or swabsticks contained inside the following Aligned Medical Solutions convenience kits with Name/Pack…