LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica HistoCore PEGASUS Tissue Processor, REF 14048858005.
Poorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks.
Read the official notice →- Date
- 2024-06-12
- Country
- United States
- Brand
- Leica Biosystems Nussloch
- Product
- Leica HistoCore PEGASUS Tissue Processor, REF 14048858005.
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution including in the states of AK, AL, AZ,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Leica Biosystems Nussloch recalls
- 2025-03-05
US
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.01
Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.
FDA · DeviceLEICA BIOSYSTEMS NUSSLOCH GMBHWorldwide - US Nationwide distribution in the states of Idaho and Ohio. The…Official notice →- 2024-10-23
US
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2;…
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause…
- 2024-08-14
US
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station.
An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
FDA · DeviceLEICA BIOSYSTEMS NUSSLOCH GMBHUS Nationwide, including government distribution.Official notice →- 2024-05-01
US
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica HistoCore PEGASUS (PROCESSOR, TISSUE, AUTOMATED)
There is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS / HistoCore PEGASUS Plus resulting from an incorrect assignment of the carryover value set for created or…
Other drugs & medical recalls in the United States
- 2024-06-12
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration…
The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide and the country of Columbia.Official notice →US
Atrium Medical Corporation recalls ADVANTA VXT, 6X80, 1GDS, FH, STR-TW. Single-Ended Slider GDS. Vascular graft
Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.
FDA · DeviceAtrium Medical CorporationWorldwide distribution - US Nationwide including Puerto Rico, and the…Official notice →US
Philips North America Llc recalls HA FlexTrak II-Patient transport functionality to transport the patient from the preparation area to the MR system, Product Number:…
Oil may leak from the HA FlexTrack Trolley onto the floor when pressing the hydraulic pedal, resulting in a potential slipping/falling hazard
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Newport Corp recalls Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply…
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when…
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized…
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the state of Virginia and the…Official notice →US
BD SWITZERLAND SARL recalls Alaris Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSite Y Sites REF 10013072 Tubing set used for delivery of…
Due to infusion set tubing drip chamber becoming detached from the tubing potentially causing delays in treatment, or unintentionally exposing users to harsh infusates.
FDA · DeviceBD SWITZERLAND SARLWorldwide - US Nationwide distribution in the states of AR, CA, CO, IL, KS,…Official notice →