Drugs & medicalClass II

Karl Storz Endoscopy recalls Karl Storz SE & CO. KG, REF 115400S, Optical Scissor, CE

Inadequate reprocessing validation evidence

Read the official notice →
Date
2024-04-10
Country
United States
Brand
Karl Storz Endoscopy
Product
Karl Storz SE & CO. KG, REF 115400S, Optical Scissor, CE
Classification
Class II
Units
US Nationwide distribution in the states of AR, CA, CO, CT, DC, FL, GA, IA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Karl Storz Endoscopy recalls

  1. 2024-11-20
  2. US
    Drugs & medicalClass II

    Karl Storz Endoscopy recalls Irrigation Tubing for KARL STORZ Endomat Select UP210 REF 031523-10 Tubing Set, Irrigation, PC

    Due to improper labeling of products. An intended use on the label has not been reviewed and approved by FDA for distribution in the U.S.

    FDA · DeviceKarl Storz EndoscopyU.S. Nationwide distribution in the states of CO, CT, DC, MD, MI, MN, NC, NY,…Official notice →
  3. 2024-10-02
  4. US
    Drugs & medicalClass II

    Karl Storz Endoscopy recalls Karl Storz - Endoskope , REF: 27092AMA, Hopkins Telescope 6, Rx only, CE 0123

    Various medical device product IFUs contain reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.

    FDA · DeviceKarl Storz EndoscopyUS Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT,…Official notice →
  5. 2024-05-22
  6. US
    Drugs & medicalClass II

    Karl Storz Endoscopy recalls LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000

    Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then…

    FDA · DeviceKarl Storz EndoscopyWorldwide distribution - US Nationwide in the states of OH, CA, FL, MA, TX,…Official notice →
  7. 2023-12-27
  8. US
    Drugs & medicalClass II

    Karl Storz Endoscopy recalls Injection Needle, 3 Fr., REF: 27030N; Injection Cannula, 6 Fr., REF: 27182A

    The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.

    FDA · DeviceKarl Storz EndoscopyUS: MI, NH, SC, PA, FL, NY, NC, MS, MO, CA, NJ, IL, NV, OK, CO, OH, WA, WI,…Official notice →
  9. 2023-09-27
  10. US
    Drugs & medicalClass II

    Karl Storz Endoscopy recalls Five S 5 3x65, REF: 0915612-06, sterile for single use, Sterile EO, Rx Only

    Flexible intubation endoscopes sterility assurance can not be confirmed.

    FDA · DeviceKarl Storz EndoscopyWorldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →
  11. 2023-06-28
  12. US
    Drugs & medicalClass II

    Karl Storz Endoscopy recalls Metal Tray, 178 x 135 x 37 mm, REF: 11580A; Metal Tray, 275 x 178 x 35 mm, REF: 11580B; Metal Tray, 373 x 178 x 35 mm, REF: 11580C; Metal…

    Metal sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the first time specific to the tray models, that also includes…

    FDA · DeviceKarl Storz EndoscopyWorldwide - US Nationwide distribution including in the states of IN, MS, MN,…Official notice →
All 17 Karl Storz Endoscopy recalls →

Other drugs & medical recalls in the United States

  1. 2024-04-10
  2. US
    Drugs & medicalClass II

    Wright Medical Technology, Inc. recalls stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty

    Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument…

    FDA · DeviceWright Medical Technology, Inc.US: TX, NE, and AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, FRANCE,…Official notice →
  3. US
    Drugs & medicalClass II

    Abbott Laboratories recalls ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.

    The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis…

    FDA · DeviceAbbott LaboratoriesWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL,…Official notice →
  4. US
    Drugs & medicalClass II

    3M Company - Health Care Business recalls 3M Tube Securement Device, Small, 1.7 IN/PO x 2.4 IN/PO, REF 1500U, Sterile; and 3M Tube Securement Device, Large, 2.2 IN/PO x…

    Manufacturing nonconformities are in some lots of the Tube Securement Device, such as the top film layer may be partially or completely missing and some devices may exhibit the appearance of an incorrect shape due to…

    FDA · Device3M Company - Health Care BusinessWorldwide distribution - US Nationwide distribution in the states of CA, CO,…Official notice →
  5. US
    Drugs & medicalClass II

    Covidien, LP recalls Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BT

    The potential for trocar seal disengagement when using mesh products incorrectly with the device.

    FDA · DeviceCovidien, LPWorldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →
  6. US
    Drugs & medicalClass II

    Coloplast Manufacturing US, LLC recalls Ureteral dilator, REF RBD014

    A possible sterility issue (e.g. a breach in sterile packaging) was detected in Coloplast's facility on some ureteral dilator products. This issue on the ureteral dilator packaging has been identified during testing in…

    FDA · DeviceColoplast Manufacturing US, LLCUS: GA, MS, TN, MN, AZ,N MD, LA, KS, WAOfficial notice →
  7. US
    Drugs & medicalClass II

    Agfa N.V. recalls Agfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality, consisting of X-Ray console, generator, tube,…

    Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is…

    FDA · DeviceAgfa N.V.Worldwide distribution - US Nationwide and the countries of Canada, Argentina,…Official notice →
See all drugs & medical recalls →