Drugs & medicalClass II

Johnson & Johnson Surgical Vision, Inc. recalls Veritas Advanced Infusion Packs. Used in ophthalmic procedures.

As a result of returned product complaints for Veritas Phaco packs related to broken or cracked irrigation luer due to a manufacturing issue.

Read the official notice →
Date
2022-10-19
Country
United States
Brand
Johnson & Johnson Surgical Vision
Product
Veritas Advanced Infusion Packs. Used in ophthalmic procedures.
Classification
Class II
Units
Worldwide Distribution. US states of AL, AR, AZ, CA, CO, DE, FL, GA, HI, IL…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Johnson & Johnson Surgical Vision recalls

  1. 2023-08-09
  2. US
    Drugs & medicalClass II

    Johnson & Johnson Surgical Vision, Inc. recalls TECNIS Symfony, Extended Range of Vision IOL, REF: ZXR00, STERILE.EO, Rx Only

    Their is a potential that intraocular lenses (IOLs) were manufactured with unintended toric fiducial marks.

    FDA · DeviceJohnson & Johnson Surgical Vision, Inc.International distribution to the countries of China, India, Korea.Official notice →
  3. US
    Drugs & medicalClass II

    Johnson & Johnson Surgical Vision, Inc. recalls VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

    Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the…

    FDA · DeviceJohnson & Johnson Surgical Vision, Inc.Worldwide - US Nationwide distribution including in the states of NC, OH, NY,…Official notice →
  4. 2023-02-01
  5. US
    Drugs & medicalClass II

    Johnson & Johnson Surgical Vision, Inc. recalls FUSION Dual Pump Packs, REF: OPO73

    Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and…

    FDA · DeviceJohnson & Johnson Surgical Vision, Inc.US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France,…Official notice →
  6. 2022-11-09
  7. US
    Drugs & medicalClass II

    Johnson & Johnson Surgical Vision, Inc. recalls THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac…

    Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation…

    FDA · DeviceJohnson & Johnson Surgical Vision, Inc.Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO,…Official notice →
  8. 2022-05-11
  9. US
    Drugs & medicalClass II

    Johnson & Johnson Surgical Vision Inc recalls TRAY RING COVER, REF OM271000, CONTAINS (1) TRAY RING COVER, STERILE R, Rx Only,

    Potential breach in the sterility barrier for tray ring covers.

    FDA · DeviceJohnson & Johnson Surgical Vision IncWorlwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →
  10. 2021-10-13
  11. US
    Drugs & medicalClass II

    Johnson & Johnson Surgical Vision Inc recalls TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation in the capsular bag. The TECNIS…

    Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification limit and the FDA recommended endotoxin specification limit.…

    FDA · DeviceJohnson & Johnson Surgical Vision IncU.S. Nationwide distribution in the state of IL.Official notice →
All 10 Johnson & Johnson Surgical Vision recalls →

Other drugs & medical recalls in the United States

  1. 2022-10-25
  2. US
    Drugs & medical

    Clorox® Recalls Pine-Sol® Scented Multi-Surface Cleaners, CloroxPro® Pine-Sol® All Purpose Cleaners, and Clorox® Professional™ Pine-Sol® Lemon Fresh Cleaners Due to…

    The recalled products may contain bacteria, including Pseudomonas aeruginosa, an environmental organism found widely in soil and water. People with weakened immune systems or external medical devices who are exposed to…

    CPSCThe Clorox Company, of Oakland, CaliforniaTo date, testing has identified bacteria in certain…Official notice →
  3. 2022-10-19
  4. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls VENCLOSE, VENCLOSE PROCEDURE PACK, 7VM, REF VC-PPH-67A. Cardiovascular procedure convenience kit.

    Impacted Venclose Procedure Packs contain a 12cm introducer/sheath component, either component #128626 - SET INTRO 7FX12CM FITS 0.018 or component #137339 - SET INTRO 6.5FX12CM FITS 0.018, instead of a 7cm…

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldCAOfficial notice →
  5. US
    Drugs & medicalClass II

    Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography System, Model: uMI 550

    The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss…

    FDA · DeviceShanghai United Imaging Healthcare Co., Ltd.Distribution to US states of TX, WI, CA, and MIOfficial notice →
  6. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Prismaflex HF20 Set (product code 109841). For use in providing continuous fluid management and renal replacement therapies.

    Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight…

    FDA · DeviceBaxter Healthcare CorporationUS NationwideOfficial notice →
  7. US
    Drugs & medicalClass II

    LUMINOSTICS, INC dba CLIP HEALTH recalls Clip COVID Rapid Antigen Test Kit (25-pack), Catalog Number: MDL002

    COVID test kit shelf life was periodically extended, as real-time stability test timepoints passed, without FDA concurrence.

    FDA · DeviceLUMINOSTICS, INC dba CLIP HEALTHUS nationwide distributionOfficial notice →
  8. US
    Drugs & medicalClass II

    Steris Corporation recalls Hu-Friedy K360 Scalpel Handle, 360 Degree Scalpel Handle, REF K360

    The scalpel holder may separate from the handle for certain lots. If the tip falls off during use, it could fall into the patient's mouth which may result in a cut to the patient's mouth and also presents a choking…

    FDA · DeviceSteris CorporationDistribution to US states of IL, MA, NC, SC, and TN, Canada, China, Japan,…Official notice →
See all drugs & medical recalls →