Drugs & medicalClass II

IsoTis OrthoBiologics, Inc. recalls OsteoCove Putty, 1.25cc, REF 56920013 and Cove Putty, 1.25cc REF 02-9200-01.

Hydration issue resulting in the product being less cohesive and moldable than intended due to an issue with the polymer.

Read the official notice →
Date
2025-01-22
Country
United States
Brand
IsoTis OrthoBiologics
Product
OsteoCove Putty, 1.25cc, REF 56920013 and Cove Putty, 1.25cc REF 02-9200-01.
Classification
Class II
Units
US Nationwide distribution in the states of AL, CA, MD. MN, MS, NC, NV, OH,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More IsoTis OrthoBiologics recalls

  1. 2026-08-05
  2. US
    Drugs & medicalClass II

    IsoTis OrthoBiologics, Inc. recalls OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095

    Demineralized bone matrix (DBM) in resorbable mesh indicated for use with autograft as a bone graft extender in the posterolateral spine and pelvis, were filled with approx. 4.5 cc DBM (fill for 45 ¿ 11 mm…

    FDA · DeviceIsoTis OrthoBiologics, Inc.US Nationwide distribution in the states of AK, NJ, TN, AR, CA, WA, OH.Official notice →
  3. 2025-01-22
  4. US
    Drugs & medicalClass II

    IsoTis OrthoBiologics, Inc. recalls OsteoSurge 100 Demineralized Bone Matrix Putty with Accell Bone Matrix, OsteoSurge 100 Demineralized Bone Matrix, 10cc, REF 56300100.

    Incorrect expiration date listed on the outer box, which does not align with the correct expiration date listed on the inner tray.

    FDA · DeviceIsoTis OrthoBiologics, Inc.US Nationwide distribution in the states of IN, TN & TX.Official notice →
All 3 IsoTis OrthoBiologics recalls →

Other drugs & medical recalls in the United States

  1. 2025-01-22
  2. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls 8MM, Small Graptor REF 470318 Is a grasping retractor Instrument used with the da Vinci X/Xi Surgical Systems in a wide variety of…

    Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments

    FDA · DeviceIntuitive Surgical, Inc.U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA,…Official notice →
  3. US
    Drugs & medicalClass II

    DRG International, Inc. recalls DRG:HYBRiD-XL Insulin. Model/Catalog Number: HYE-5361

    The user of the assay may either face a failure message during re-calibration of the Hybrid.XL instrument using the re-calibration R1 and R2 of the device or find controls out of specification (failed Control High)…

    FDA · DeviceDRG International, Inc.Worldwide - US Nationwide distribution in the state of New Jersey and the…Official notice →
  4. US
    Drugs & medicalClass II

    SIN SISTEMA DE IMPLANTE NACIONAL S.A recalls Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N

    Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the…

    FDA · DeviceSIN SISTEMA DE IMPLANTE NACIONAL S.AUS distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND…Official notice →
  5. US
    Drugs & medicalClass II

    GE Medical Systems, LLC recalls Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring…

    There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.

    FDA · DeviceGE Medical Systems, LLCWorldwide - US Nationwide distribution in the states of OH, TX and the…Official notice →
  6. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1 ) ANEURYSM PACK , Catalog Number: SCVHFAPBM1; 2 ) ANEURYSM PACK , Catalog Number: SCVHFAPBM3; 3 )…

    Potential endotoxin contamination in non-sterile surgical strips and patties in Presource kits.

    FDA · DeviceCardinal Health 200, LLCWorldwide distribution - US Nationwide and the countries of Canada, Saudi…Official notice →
  7. US
    Drugs & medicalClass II

    Hologic, Inc. recalls Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.

    Potential weak pouch seal on Panther Fusion GBS assay cartridge, causing potential for invalid or incorrect results leading to a possible delay of treatment.

    FDA · DeviceHologic, Inc.AR, CA, CT, FL, IN, KS, LA, MA, ME, MI, NC, NJ, NV, NY, PR, TX, UT, VT, WA,…Official notice →
See all drugs & medical recalls →