Drugs & medicalClass II

Intellijoint Surgical, Inc. recalls Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint…

A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may…

Read the official notice →
Date
2026-10-14
Country
United States
Brand
Intellijoint Surgical, Inc.
Product
Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP,…
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

Other drugs & medical recalls in the United States

  1. 2026-10-14
  2. US
    Drugs & medicalClass II

    Steris Corporation recalls HexaVue Focus Integration System, a single, all-in-one system that handles operating room video routing, surgical video integration, and…

    A software issue may result in unintended mismatch of patient information on manually printed images. Incorrect patient information appearing on printed images could create a documentation discrepancy or incorrect…

    FDA · DeviceSteris CorporationWorldwide - US Nationwide and the countries of Australia, Canada, New Zealand.Official notice →
  3. US
    Drugs & medicalClass II

    LeMaitre Vascular, Inc. recalls TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;

    Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.

    FDA · DeviceLeMaitre Vascular, Inc.US Nationwide distribution in the states of CA, FL, GA, KY, LA, NJ, OH, PA, SC.Official notice →
  4. US
    Drugs & medicalClass III

    Mindray DS USA, Inc. recalls ePM Series Patient Monitor. Patient Monitor Series: ePM 10M/ePM 12M/ePM 12MA/ ePM 15M/ePM 10A

    Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions

    FDA · DeviceMindray DS USA, Inc.Worldwide - Nationwide and the country of Canada.Official notice →
  5. US
    Drugs & medicalClass II

    Canon Medical System, USA, INC. recalls Alphenix INFX-8000C/B5

    Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error

    FDA · DeviceCanon Medical System, USA, INC.US Nationwide distribution in the state of ID.Official notice →
  6. US
    Drugs & medicalClass II

    POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC recalls SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;

    During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting…

    FDA · DevicePOLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLCWorldwide - US Nationwide distribution in the states and territory of AK, AR,…Official notice →
  7. US
    Drugs & medicalClass II

    AVID Medical, Inc. recalls ROBOTIC PROSTATECTOMY PACK; Model/Item Number (Kit): NAMC423-11;

    Product may be potentially contaminated with a lubricant used during the manufacturing process.

    FDA · DeviceAVID Medical, Inc.US Nationwide distribution in the states of AZ, CA, FL, MD, NE, TX.Official notice →