Drugs & medicalClass II

Integrity Implants Inc. recalls LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal…

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

Read the official notice →
Date
2022-12-14
Country
United States
Brand
Integrity Implants
Product
LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal…
Classification
Class II
Units
Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Integrity Implants recalls

  1. 2025-01-29
  2. US
    Drugs & medicalClass II

    Integrity Implants Inc. recalls Brand Name: N/A Product Name: Repeater Bone Tamp Model/Catalog Number: ASY-00132 Software Version: N/A Product Description: The Repeater…

    Due to incompatibility of bone tamp (Rev A) and bone funnels (Rev C).

    FDA · DeviceIntegrity Implants Inc.U.S. Nationwide distribution in the states of CA, F, GA, KY, LA, MA, MD, TN…Official notice →
  3. 2022-12-14
  4. US
    Drugs & medicalClass II

    Integrity Implants Inc. recalls LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545.…

    Potential for intraoperative screw head separation at weld location.

    FDA · DeviceIntegrity Implants Inc.US distribution to California, Indiana, New York, Oklahoma, Washington DCOfficial notice →
All 3 Integrity Implants recalls →

Other drugs & medical recalls in the United States

  1. 2022-12-14
  2. US
    Drugs & medicalClass II

    Insulet Corporation recalls Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor,…

    The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the…

    FDA · DeviceInsulet CorporationDomestic distribution nationwide.Official notice →
  3. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

    The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power…

    FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.Official notice →
  4. US
    Drugs & medicalClass II

    Integrity Implants Inc. recalls LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545.…

    Potential for intraoperative screw head separation at weld location.

    FDA · DeviceIntegrity Implants Inc.US distribution to California, Indiana, New York, Oklahoma, Washington DCOfficial notice →
  5. US
    Drugs & medicalClass II

    Carefusion 2200 Inc recalls CareFusion V. Mueller TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS, 0.7MM TIP OVERALL LENGTH 9-1/2" (240MM) , REF F-5008

    XXX

    FDA · DeviceCarefusion 2200 IncNationwide, Austria, Canada, Egypt, Germany, Hong Kong, Israel, Philippines,…Official notice →
  6. US
    Drugs & medicalClass II

    Carefusion 2200 Inc recalls Carefusion V. Mueller" CUSHING BAYONET BIPOLAR FORCEPS, 2.0MM TIP, INSULATED OVERALL LENGTH 7-1/2" (190MM), REF F-2004

    BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power…

    FDA · DeviceCarefusion 2200 IncNationwide, Austria, Canada, Egypt, Germany, Hong Kong, Israel, Philippines,…Official notice →
  7. US
    Drugs & medicalClass II

    Medacta Usa Inc recalls NextAR Shoulder Surgical Technique, Shoulder NextAR RSA Platform, CT based computer-assisted surgical navigation platform used to perform…

    A warning has been added to the Surgical Technique for NextAR Shoulder, advising verification of suitable patient conditions prior to placement of the NextAR Shoulder Target Holder: WARNING Target fixation requires…

    FDA · DeviceMedacta Usa IncAL, FL, NH, CO, IL, SC, AROfficial notice →
See all drugs & medical recalls →