Drugs & medicalClass III

Instrumentation Laboratory recalls HemosIL SynthAFax. Partial Thromboplastin Time Tests.

Recalled lots were manufactured with double the amount of preservative concentration.

Read the official notice →
Date
2025-12-24
Country
United States
Brand
Instrumentation Laboratory
Product
HemosIL SynthAFax. Partial Thromboplastin Time Tests.
Classification
Class III
Units
Domestic: CA, CO, FL, NY; Foreign: AR, AT, CA, CZ, DE, DK, ES, FR, GB, IE, IL,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Instrumentation Laboratory recalls

  1. 2026-08-05
  2. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system

    Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

    FDA · DeviceInstrumentation LaboratoryUS Nationwide distribution.Official notice →
  3. 2026-03-18
  4. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls GEM Premier 5000 PAK, Part No. 00055360004.

    Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including…

    FDA · DeviceInstrumentation LaboratoryWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  5. 2026-01-21
  6. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls The product is used for the evaluation of the intrinsic coagulation pathway, APTT substitution test and the monitoring of heparin…

    Potential for microbial contamination.

    FDA · DeviceInstrumentation LaboratoryWorldwide - US Nationwide distribution in the states of CA, FL, IL, MN, MS,…Official notice →
  7. 2025-08-20
  8. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls HemosIL LMW Heparin Controls; Part Number: 0020300200;

    Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL…

    FDA · DeviceInstrumentation LaboratoryWorldwide - US Nationwide and the countries of Algeria, Andorra, Argentina,…Official notice →
  9. 2025-02-05
  10. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic

    Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range.

    FDA · DeviceInstrumentation LaboratoryWorldwide - US Nationwide and the countries of Andorra, Algeria, Armenia,…Official notice →
  11. 2023-11-29
  12. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007

    Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled…

    FDA · DeviceInstrumentation LaboratoryUS States: MA, MN, NY. India.Official notice →
All 14 Instrumentation Laboratory recalls →

Other drugs & medical recalls in the United States

  1. 2025-12-24
  2. US
    Drugs & medicalClass II

    Copan Italia recalls 1ml of DTT in liquid phase, in PET tube with capture cap, in bulk + sputum transfer device, for WASP* *Compatible with Copan s WASP…

    Treatment reagent functionality in respiratory specimens liquification cannot be guaranteed.

    FDA · DeviceCopan ItaliaUS Distribution to states of: CA, CO, FL, GA, MN, MS, NH, NJ, NY, OH, PA, TN,…Official notice →
  3. US
    Drugs & medicalClass II

    DICOM Grid, Inc. recalls Intelerad InteleShare software, with ProViewer component

    Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare viewer; Multiplanar reconstruction…

    FDA · DeviceDICOM Grid, Inc.US: AZ, TX, GA, NV, MO, MI, IL, OH, CT, OR, FL, NY, NM, MN, PA, AK, ND, NJ,…Official notice →
  4. US
    Drugs & medicalClass II

    PATH recalls MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)

    Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can…

    FDA · DevicePATHDomestic: WI. International: Ireland.Official notice →
  5. US
    Drugs & medicalClass II

    SEASPINE ORTHOPEDICS CORPORATION recalls Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation…

    The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However,…

    FDA · DeviceSEASPINE ORTHOPEDICS CORPORATIONUS: CT, TX, NV, CO, CA, WA, NJ, OR, IN, MI, FL, PA, IL, VA, AL, RI, MO, AZ,…Official notice →
  6. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls 5008X CAREsystem +CLiC +CDX;

    Several software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if certain conditions are fulfilled.

    FDA · DeviceFresenius Medical Care Holdings, Inc.Domestic: LA, MA, MN;Official notice →
  7. US
    Drugs & medicalClass II

    GE Healthcare GmbH recalls GE HealthCare ViewPoint 6, Catalog Number H47601AA; Radiological image processing system

    GE HealthCare has become aware that the Patient History tab in ViewPoint 6 versions 6.15.3 and 6.15.4 ultrasound software may not include complete patient medical history information.

    FDA · DeviceGE Healthcare GmbHWorldwideOfficial notice →
See all drugs & medical recalls →