IHB OPERATIONS B.V. recalls Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware…
Read the official notice →- Date
- 2025-04-30
- Country
- United States
- Brand
- Ihb Operations
- Product
- Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
- Classification
- Class II
- Units
- US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Ihb Operations recalls
- 2023-12-13
US
IHB OPERATIONS B.V. recalls Freedom Boom models series: F110, F120, F130, F320, F330, F350, F410, and F420 with built in 42A Lightweight Spring Arm part numbers…
Weakened connection point on the adjustable suspension arm, potentially causing damage or a break.
FDA · DeviceIHB OPERATIONS B.V.US: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MB, MD, ME, MI,…Official notice →
Other drugs & medical recalls in the United States
- 2025-04-30
US
Remote Diagnostic Technologies Ltd. recalls Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007…
software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide - US Nationwide distribution including in the states of AL, AZ, CA,…Official notice →US
Abbott Molecular, Inc. recalls The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001,…
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the…
US
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer
When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose…
US
Philips North America Llc recalls Philips Brilliance CT Big Bore, Software Version 4.8.0.10430. Computed Tomography X-Ray System.
Unintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.
FDA · DevicePhilips North America LlcWorldwide distribution: US (Nationwide); and OUS (Foreign) to countries of:…Official notice →US
Radiometer Medical ApS recalls Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90 Series - Radiometer ABL90 FLEX PLUS Model Number…
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software…
FDA · DeviceRadiometer Medical ApSU.S.: AL, AR, AZ, BM CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →US
Edge Biologicals Inc recalls Edge Biologicals STERILE WATER, 3ML, 100/PK CAT T-0780 Sterile Water, 3mL, 10/PK - Sterile water is used in making dilutions of…
Due to product outer packaging incorrectly labeled.