Drugs & medicalClass II

Howmedica Osteonics Corp. recalls 1. EXETER V40 STEM 44MM NO 2, Model/Catalog Number: 0580-1-442; 2. EXETER V40 STEM 37.5MM NO 0, Model/Catalog Number: 0580-1-352;

a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted.

Read the official notice →
Date
2025-10-08
Country
United States
Brand
Howmedica Osteonics
Product
1. EXETER V40 STEM 44MM NO 2, Model/Catalog Number: 0580-1-442; 2. EXETER V40 STEM 37.5MM NO 0, Model/Catalog Number: 0580-1-352;
Classification
Class II
Units
International distribution in the country of United Kingdom.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Howmedica Osteonics recalls

  1. 2026-06-24
  2. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.

    Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot…

    FDA · DeviceHowmedica Osteonics Corp.US Nationwide distribution in the states of AZ, CA, GA, MI, NC, NJ, NY, WV.Official notice →
  3. 2026-03-25
  4. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

    A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and…

    FDA · DeviceHowmedica Osteonics Corp.Domestic: Nationwide Distribution; International: Australia, Germany, Belgium,…Official notice →
  5. 2025-12-31
  6. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR)…

    When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

    FDA · DeviceHowmedica Osteonics Corp.Worldwide distribution - US Nationwide and the countries of Canada, Germany,…Official notice →
  7. 2024-10-16
  8. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls MAKO Integrated Cutting System (MICS) HANDPIECE, Catalog Number 209063

    Stryker has identified an issue with the characterization process that impacted the constant values assigned to the associated MICS Handpieces.

    FDA · DeviceHowmedica Osteonics Corp.Worldwide distribution - US Nationwide and the countries of Australia, Germany.Official notice →
  9. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications during primary and…

    Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.

    FDA · DeviceHowmedica Osteonics Corp.US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI.…Official notice →
  10. 2024-09-04
  11. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls Custom Made Device - GMRS EXTENSION Piece 40MM LEFT, 35 Deg EXT Rotated Tabs (Patient Specific); Catalog Number: C-MM100434-00;

    A patient specific, custom-made device was supplied for a surgical procedure that did not match the design proposal.

    FDA · DeviceHowmedica Osteonics Corp.International distribution to the country of Australia.Official notice →
All 28 Howmedica Osteonics recalls →

Other drugs & medical recalls in the United States

  1. 2025-10-08
  2. US
    Drugs & medicalClass II

    bioMerieux, Inc. recalls VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system consisting of…

    Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

    FDA · DevicebioMerieux, Inc.International distribution in the countries of Albania, Armenia, Australia,…Official notice →
  3. US
    Drugs & medicalClass I

    Medline Industries, LP recalls Medline Kits: 1) VALVE PACK-LF, Model Number: DYNJ0415366P; 2) VALVE PACK-LF, Model Number: DYNJ0415366Q

    The kits contain certain lots of cannula products where the catheter may not retain its shape.

    FDA · DeviceMedline Industries, LPUS Nationwide distribution in the state of CA.Official notice →
  4. US
    Drugs & medicalClass I

    Siemens Medical Solutions USA, Inc recalls MAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253, K141977, K142515,…

    There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a…

    FDA · DeviceSiemens Medical Solutions USA, IncWorldwide distribution - US Nationwide and the countries of United Arab…Official notice →
  5. US
    Drugs & medicalClass II

    Zap Surgical Systems recalls ZAP-X Radiosurgery System, Model: 300150

    If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software…

    FDA · DeviceZap Surgical SystemsWorldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the…Official notice →
  6. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF…

    Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use.…

    FDA · DeviceBard Peripheral Vascular IncWorldwide - US Nationwide distribution including in the states of NJ, NY, FL,…Official notice →
  7. US
    Drugs & medicalClass II

    North American Rescue LLC. recalls JUNCTIONAL EMERGENCY TREATMENT TOOL (JETT), REF: 30-0088, a component inside kits: KIT, TCCC Training- Complete (Modules 1-3), REF:…

    Device used for unilateral or bilateral occlusion of blood flow to the lower limbs has a potential manufacturing defect that may compromise the structural integrity of the device and device performance during use.

    FDA · DeviceNorth American Rescue LLC.Worldwide - US Nationwide distribution in the states of SC, OK, ID, KY, VA,…Official notice →