Drugs & medicalClass II

Greiner Bio-One North America, Inc. recalls Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322

Blood collection tubes were manufactured with the label applied too high on the tube, which shifted the printed fill level mark upward, which could result in unnecessary sample rejection and repeat collections.

Read the official notice →
Date
2025-11-05
Country
United States
Brand
Greiner Bio-One
Product
Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322
Classification
Class II
Units
US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Greiner Bio-One recalls

  1. 2025-11-26
  2. US
    Drugs & medicalClass II

    Greiner Bio-One North America, Inc. recalls Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product…

    Some of the Vacuette¿ CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat…

    FDA · DeviceGreiner Bio-One North America, Inc.U.S.: AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS,…Official notice →
  3. 2025-01-15
  4. US
    Drugs & medicalClass II

    Greiner Bio-One GmbH recalls VACUETTE SAFELINK, REF: 450210

    The VACUETTE SAFELINK Holder with male luer lock may leak blood due to the protective rubber sleeve bunching up and not retracting when using the device.

    FDA · DeviceGreiner Bio-One GmbHUS Nationwide distribution in the states of IL, PA.Official notice →
  5. 2024-07-24
  6. US
    Drugs & medicalClass II

    Greiner Bio-One North America, Inc. recalls VACUETTE¿ TUBE 8 ml CAT Serum Separator Clot Activator 16x100 red cap-yellow ring, non-ridged

    The firm received reports that some blood collection tubes have experienced issues with clotting and poor separation due to missing additive. The failure of the device would likely require sample recollection which…

    FDA · DeviceGreiner Bio-One North America, Inc.Nationwide Distribution to states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA,…Official notice →
  7. 2024-01-24
  8. US
    Drugs & medicalClass II

    Greiner Bio-One North America, Inc. recalls VACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate, Item #454297, 13x75 grey cap-black ring, non-ridged.

    Tubes have missing additive/anticoagulant resulting in clotting in most cases. A completely clotted sample would lead to no results being reported, which could lead to a delay in treatment or diagnosis. An erroneous…

    FDA · DeviceGreiner Bio-One North America, Inc.US Nationwide distribution in the states of CT, FL, IA, IL, MO, NC, NY, PA,…Official notice →
  9. 2023-05-24
  10. US
    Drugs & medicalClass II

    Greiner Bio-One North America, Inc. recalls greiner bio-one VACUETTE TUBE 6 ml K3E K3EDTA, 13x100 pink cap-black ring, non-ridged, 24 racks of 50 pcs., 1200 pcs in…

    Some of the tubes may be incorrectly labeled.

    FDA · DeviceGreiner Bio-One North America, Inc.US Nationwide distribution in the states of WI and GA.Official notice →
  11. 2022-01-26
  12. US
    Drugs & medicalClass II

    Greiner Bio-One North America, Inc. recalls greiner bio-one VACUETTE TUBE, 5 ml LH Lithium Heparin Sep, 13x100 green cap-yellow ring, non-ridged, 24 racks of 50 pcs.,…

    Some of the blood collection tubes show clotting due to variation of Lithium Heparin.

    FDA · DeviceGreiner Bio-One North America, Inc.Distribution was made to FL, IL, MO, NC, NM, NY, PA, TN, TX, and VA. There was…Official notice →
All 8 Greiner Bio-One recalls →

Other drugs & medical recalls in the United States

  1. 2025-11-05
  2. US
    Drugs & medicalClass II

    MicroSurgical Technology Inc recalls MST ArcDUO 9mm, REF: ARC-0002, single-use, sterile ophthalmic knife, used with Suction Set, Part: 48-1000-01

    An ophthalmic knife has a suction set (not direct patient-contacting), which contains springs contaminated with an unidentified residue that may potentially cause eye irritation

    FDA · DeviceMicroSurgical Technology IncUS Nationwide distribution in the state of GA.Official notice →
  3. US
    Drugs & medicalClass II

    Medistim Asa recalls Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011…

    Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.

    FDA · DeviceMedistim AsaDomestic distribution to CO, MI, MN, TN, TX. International distribution to…Official notice →
  4. US
    Drugs & medicalClass II

    Vantive US Healthcare LLC recalls Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and…

    Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using…

    FDA · DeviceVantive US Healthcare LLCWorldwide distribution - US Nationwide and the countries of Australia, Hong…Official notice →
  5. US
    Drugs & medicalClass II

    DAYE (ANNE'S DAY LTD) recalls Refill 3 Month All in One Bundle; Model Number: FG-BNDL-PC-MM-REFAIO;

    Product lacks 510(k) clearance.

    FDA · DeviceDAYE (ANNE'S DAY LTD)US Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.

    Potential issue that can cause the device to register "phantom" touches in the lower-left corner of the touchscreen during operation.

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution in the state of Minnesota.Official notice →
  7. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS recalls Model: Incisive CT for Brazil SKD; Product Code (REF): 728146; Software Versions: 5.1.0.X & 5.1.1.X;

    Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/…

    FDA · DevicePHILIPS MEDICAL SYSTEMSDomestic: AL, AR, CA, CT, FL, GA, IL, IN, KY, MD, MI, OH, PA, PR, TN, TX, UT,…Official notice →
See all drugs & medical recalls →