Getinge Usa Sales Inc recalls Flow-c Anesthesia System : intended for use in administering anesthesia while controlling the entire ventilation of patient Product…
Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being…
Read the official notice →- Date
- 2022-06-29
- Country
- United States
- Brand
- Getinge Usa Sales
- Product
- Flow-c Anesthesia System : intended for use in administering anesthesia while controlling the entire ventilation of patient Product…
- Classification
- Class I
- Units
- US Distribution in states of Colorado, Florida, New Jersey, Indiana, Missouri,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Getinge Usa Sales recalls
- 2022-08-31
US
Getinge Usa Sales Inc recalls Servo-air Ventilator System-The Servo-air Ventilator System is: Intended for respiratory support, monitoring and treatment of pediatric…
Device does not meet regulatory requirements of stability while stationary, which requires mechanical equipment, other than fixed mechanical equipment, that is intended to be used on the floor or on a table shall be…
FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide and the countries of Australia (AU),…Official notice →US
Getinge Usa Sales Inc recalls Maquet Critical Care AB/Getinge Servo-u Ventilator System, Part Number 6694800 Servo-n Ventilator System, Part Number 6688600 Servo-u MR…
Ventilators may generate a combination of alarms which may result in loss of communication, technical alarms being triggered, and loss of ventilation. If the device stops ventilating due to this issue, it may lead to…
FDA · DeviceGetinge Usa Sales IncWorldwide distribution. US nationwide, Poland, Spain, France, Germany, Czech…Official notice →- 2022-07-06
US
Getinge Usa Sales Inc recalls VOLISTA StandOP Surgical Light, Product Code/Part Numbers ARD568811901 VLT600SF AIM STP ARD568811911 VLT600SF AIM STP ARD568811961…
Paint particles may detach and fall in the operating field. In the worst case scenario, infection, organ irritation, hypersensitivity reaction, and tissue granuloma may result if paint particles fall during surgical…
FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide distribution in the states of AR, AZ,…Official notice →- 2022-06-15
US
Getinge Usa Sales Inc recalls Getinge CM320 WUWD Series Washer Disinfector-For use as a multi-chamber washer-disinfector for the washing, disinfecting and drying…
Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a…
FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution in the states of MD, MN, ND, NE, NY, PA, VA.Official notice →- 2022-05-25
US
Getinge Usa Sales Inc recalls Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW.
Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and hazardous situations.
FDA · DeviceGetinge Usa Sales IncWorldwide distribution: (US) AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KY, LA,…Official notice →- 2022-01-19
US
Getinge Usa Sales Inc recalls Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia…
A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.
FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide distribution in the states of VA and…Official notice →
Other drugs & medical recalls in the United States
- 2022-06-29
US
Maquet Cardiovascular, LLC recalls INTERGARD Woven Vascular Graft is a woven polyester graft coated with a bovine collagen. It has an external velour surface and…
Mislabeled: Product labeled as Intergard Woven Straight graft REF IGW0032-30, D: 8 mm, L: 30 cm, contained an Intergard Woven Straight graft of D: 32 mm and L: 30 cm
FDA · DeviceMaquet Cardiovascular, LLCUS Nationwide distribution in the state of SC.Official notice →US
Beaver Visitec International, Inc. recalls BVI Wet -Field Eraser, 18GA Blunt Tip, 45Deg.-Intended to remove tissue and control bleeding by the use of high frequency…
Boxes labeled as Wet-Field Eraser, 18GA Blunt Tip, 45Deg., were shipped containing incorrect product a Wet-Field Eraser, 18GA Bevel Tip, Straight, Part Number: 221250
FDA · DeviceBeaver Visitec International, Inc.US Nationwide distribution in the states of AZ, CA, CT, DC, FL, GA, IN, MA,…Official notice →US
New Standard Device Inc recalls Metalogix Slotted 90 Tubular Wrench x 10mm REF 900104 LOT A29618A
Due to manufacturing processing error, wrench causing stripping when torqueing nuts.
US
Olympus Corporation of the Americas recalls EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus,…
Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AL, FL, NC, NY.Official notice →US
Olympus Corporation of the Americas recalls Colonovideoscope Model CF-H180AL
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the state of KY.Official notice →US
Bard Peripheral Vascular Inc recalls REF: PIG1260K, 6Fr, X10, Safe-t-Centesis Catheter drainage Kit, Sterile EO,
There is the potential that a 6 Fr Catheter Drainage kit contains an 8 Fr needle instead of a 6 Fr needle. They can be used in conjunction with the following drainage methods: Universal extension set and drainage bag,…
FDA · DeviceBard Peripheral Vascular IncWorldwide distribution - US Nationwide distribution in the states of MS, NC,…Official notice →