Drugs & medicalClass II

GE Healthcare, LLC recalls Signa 1.5T TwinSpeed Magnetic Resonance Imaging System

There is potential for the images to be flipped left to right.

Read the official notice →
Date
2022-04-13
Country
United States
Brand
GE Healthcare
Product
Signa 1.5T TwinSpeed Magnetic Resonance Imaging System
Classification
Class II
Units
Worldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More GE Healthcare recalls

  1. 2022-06-15
  2. US
    Drugs & medicalClass I

    GE Healthcare, LLC recalls Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstr¿om Carestation and Engstr¿om PRO…

    Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.

    FDA · DeviceGE Healthcare, LLCWorldwide Distribution: US (nationwide) to AL, AR, AZ, CA, CO, CT, FL, GA, HI,…Official notice →
  3. 2022-04-27
  4. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls GE Healthcare MR Systems, Magnetic Resonance imaging (MRI). SIGNA Premier, SIGNA Architect, SIGNA Pioneer, Discovery MR750w 3.0T,…

    During the de-installation of an MR system magnet, if all rigging and hardware that attach to the magnet for transportation are not properly installed and secured, the magnet may fall, leading to potential injury.

    FDA · DeviceGE Healthcare, LLCWorldwide Distribution: US (nationwide), including Puerto Rico and the U.S.…Official notice →
  5. 2022-04-20
  6. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls MUSE Cardiology Information System, Model Numbers: a) 3101011-040, b) 3101010-401, c) 3101010-009, d) 3101010-008

    Two scenarios may cause edits to measurements and diagnosis statements can be lost after a test is Signed in the MUSE NX web client.

    FDA · DeviceGE Healthcare, LLCUS and Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, England,…Official notice →
  7. 2022-04-13
  8. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls Revolution Apex, Revolution CT with Apex Edition

    The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.

    FDA · DeviceGE Healthcare, LLCWorldwide - US Nationwide distribution.Official notice →
  9. 2022-03-23
  10. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls GE Healthcare CARESCAPE Central Station (CSCS) V2.0x MAI700 (Integrated), REF 2082278-001, and MAS700 (Desktop, REF 2082279-001.

    If the CARESCAPE Central Station v2.0 is used with an unapproved keyboard, the audio can be muted resulting in loss of audible alarms.

    FDA · DeviceGE Healthcare, LLCWorldwide distribution - US nationwide, including Puerto Rico. There was also…Official notice →
  11. 2022-02-16
  12. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Artist system, nuclear magnetic resonance imaging system.

    The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.

    FDA · DeviceGE Healthcare, LLCDistribution was nationwide, including Puerto Rico and the Virgin Islands.…Official notice →
All 41 GE Healthcare recalls →

Other drugs & medical recalls in the United States

  1. 2022-04-13
  2. US
    Drugs & medicalClass II

    Shimadzu Medical Systems recalls MODEL: X-RAY TV SYSSTEM SONIALVISION safire17

    It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

    FDA · DeviceShimadzu Medical SystemsU.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN,…Official notice →
  3. US
    Drugs & medicalClass II

    Stradis Medical, LLC dba Stradis Healthcare recalls Coryell Memorial Hosp Total, Joint Pack - Surgical Kit

    The outer bag may be perforated or torn, which may result in a breach in the sterility of the kit.

    FDA · DeviceStradis Medical, LLC dba Stradis HealthcareUS Nationwide distribution in the states of IN, TX, and PA.Official notice →
  4. US
    Drugs & medicalClass II

    Remote Diagnostic Technologies Ltd. recalls Tempus LS, Type DGE1, CAT 00-3010, REF 3.940590, Software version 1.3.5, UDI: (01)07613365001693

    Distribution of Defibrillators that are not approved or cleared for distribution US Market.

    FDA · DeviceRemote Diagnostic Technologies Ltd.US Nationwide distribution in the states of IL and KS.Official notice →
  5. US
    Drugs & medicalClass II

    SA3, LLC recalls Silatrix Oral Gel Polymerized Sucralfate Gel 10% (1 gm/10 gm

    Due to failure of microbial testing for total aerobic microbial count (TAMC).

    FDA · DeviceSA3, LLCU.S. Nationwide distribution in the states of CA, FL and GA.Official notice →
  6. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls Revolution Apex, Revolution CT with Apex Edition

    The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.

    FDA · DeviceGE Healthcare, LLCWorldwide - US Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Philips Healthcare recalls Philips StentBoost Live R2.0 application, when used with the Philips Azurion with software release R2.x

    Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system,…

    FDA · DevicePhilips HealthcareUS Nationwide distribution.Official notice →
See all drugs & medical recalls →