Fresenius Medical Care Holdings, Inc. recalls 3ML Syringe Luer Lock with needle 20GX1, Part Number 15-03ML-0, 100 per box, 24 boxes per case
There is a potential for blood backfill leakage and foreign material.
Read the official notice →- Date
- 2023-10-18
- Country
- United States
- Brand
- Fresenius Medical Care Holdings
- Product
- 3ML Syringe Luer Lock with needle 20GX1, Part Number 15-03ML-0, 100 per box, 24 boxes per case
- Classification
- Class I
- Units
- US Nationwide.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Fresenius Medical Care Holdings recalls
- 2024-01-03
US
Fresenius Medical Care Holdings, Inc. recalls Liberty Select Cycler, Model numbers RTLR180343 and RTLR180111
Affected Liberty Select cyclers may display an incorrect time stamp on the treatment data report sent to the clinic.
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AL, CA, FL, IL, IN, LA, MI, MO,…Official notice →- 2023-12-13
US
Fresenius Medical Care Holdings, Inc. recalls Stay Safe Cap
Updating the labeling to include additional warnings and contraindications for potential side effects.
- 2023-10-18
US
Fresenius Medical Care Holdings, Inc. recalls 2008T HD SYS. W/O CDX BLUESTAR
Potential PCBA leaching from tubing of hemodialysis machines.
FDA · DeviceFresenius Medical Care Holdings, Inc.Domestic: Nationwide Distribution.Official notice →- 2023-06-07
US
Fresenius Medical Care Holdings, Inc. recalls Sodium Chloride 0.9% Injection, USP, 1000 mL bags, 12 bags per carton, Rx only, Fresenius Medical Care North America,…
Lack of assurance of sterility: The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging.
FDA · DrugFresenius Medical Care Holdings, Inc.Product was distributed to 32 clinics in SC, NC and VA.Official notice →- 2022-12-28
US
Fresenius Medical Care Holdings, Inc. recalls 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare…
(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false…
- 2022-12-14
US
Fresenius Medical Care Holdings, Inc. recalls 0.9% Sodium Chloride Injection, USP, Each 100 mL contains: SODIUM CHLORIDE, USP - 900 mg, WATER FOR INJECTION, USP - qs,…
Lack of Assurance of Sterility: Leakage of 0.9% Sodium Chloride for Injection, 1L, 12pk Saline Solution.
FDA · DrugFresenius Medical Care Holdings, Inc.Product was distributed Nationwide in the USA.Official notice →
Other drugs & medical recalls in the United States
- 2023-10-18
US
Flower Orthopedics Corporation recalls Coventus Flower Orthopedics Flex-Thread TM 3.2mm Cannulated Reamer-used to ream and subsequently open the intramedullary canal to…
Device is breaking while in use, potential for patients to require a revision surgery.
US
LivaNova Deutschland GmbH recalls 3T Heater Cooler System, REF: 16-02-85
The firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device.
FDA · DeviceLivaNova Deutschland GmbHUS Nationwide distribution in the states of CA, IN, GA, NC, TX, AZ, FL, MN,…Official notice →US
Fresenius Medical Care Holdings, Inc. recalls 2008T HD SYS. W/O CDX BLUESTAR
Potential PCBA leaching from tubing of hemodialysis machines.
FDA · DeviceFresenius Medical Care Holdings, Inc.Domestic: Nationwide Distribution.Official notice →US
Scynexis, Inc. recalls BREXAFEMME (Ibrexafungerp) tablets 150 mg, 4 count cartons, Rx Only, Manufactured for and distributed by SCYNEXIS, Inc., Jersey City, NJ…
Cross Contamination with Other Products: Potential cross-contamination of ibrexafungerp citrate drug substance with ezetimibe (a non-antibacterial beta-lactam compound).
- 2023-10-11
US
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Iron_2 (IRON_2) Reagents, Material Numbers 10377510 (6 x 350 tests) and 10341118 (7 x 145 tests)
There is a potential for a positive bias on Quality Control (QC) and patient sample results when using some IRON_2 reagent wedge pairs. When the issue occurs, the observed bias will vary wedge to wedge.
US
Qiagen GmbH recalls therascreen ¿KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Rotor-Gene Q MDx instrument for the detection of seven somatic…
Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be…
FDA · DeviceQiagen GmbHWorldwide - US Nationwide distribution in the states of AZ, CA, FL, IL, IN,…Official notice →