Drugs & medicalClass II

Fresenius Kabi USA, LLC recalls LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Product Code: SET-0032-25.

Potential for external cassette leaks

Read the official notice →
Date
2025-08-06
Country
United States
Brand
Fresenius Kabi
Product
LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Product Code: SET-0032-25.
Classification
Class II
Units
US Nationwide distribution in the states of CA, CO, DE, FL, GA, ID, IL, MA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Fresenius Kabi recalls

  1. 2025-12-10
  2. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls IVENIX INFUSION SYSTEM, LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Large Volume…

    Incorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution in the states of IL, MD, MI, OK, SC, TX, VA & WI.Official notice →
  3. 2025-12-03
  4. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls Famotidine Injection, USP, 20 mg per 2 mL (10 mg per mL), 2 mL vials, Rx only, Fresenius Kabi USA, LLC, NDC 63323-739-11 (unit of use),…

    Microbial Contamination of Sterile Products; out of limit results obtained for endotoxin testing.

    FDA · DrugFresenius Kabi USA, LLCNationwide within the United States as well as AK, HI, and PR.Official notice →
  5. 2025-11-05
  6. US
    Drugs & medicalClass II

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.

    Potential issue that can cause the device to register "phantom" touches in the lower-left corner of the touchscreen during operation.

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution in the state of Minnesota.Official notice →
  7. 2025-06-18
  8. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular…

    Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.

    FDA · DeviceFresenius Kabi USA, LLCUS distribution: CA, CO, DE, GA, IL, MD, MI, MN, MS, NC, NJ, NV, OK, SC, TX,…Official notice →
  9. 2025-02-12
  10. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the…

    Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2.…

    FDA · DeviceFresenius Kabi USA, LLCUS distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI.Official notice →
  11. 2025-01-15
  12. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

    Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.

    FDA · DeviceFresenius Kabi USA, LLCUS distribution to states of: ID, MN, NV, TX, VA, WI.Official notice →
All 31 Fresenius Kabi recalls →

Other drugs & medical recalls in the United States

  1. 2025-08-13
  2. US
    Drugs & medicalClass II

    Alembic Pharmaceuticals Limited recalls Doxepin Hydrochloride Capsules, USP, 10 mg, 100 Capsules, Rx only, Manufactured by: Alembic Pharmaceuticals Limited, Panela v…

    CGMP Deviations: Presence of Nitrosamine Drug Substance Related Impurity above the proposed interim limit.

    FDA · DrugAlembic Pharmaceuticals LimitedNationwide in the USA.Official notice →
  3. US
  4. US
    Drugs & medicalClass II

    Breckenridge Pharmaceutical, Inc. recalls Duloxetine Delayed-Release Capsules, USP, 60mg, 1,000-count bottles, Rx Only, Mfr. by: Towa Pharmaceutical Europe, S.L.…

    CGMP Deviations: Presence of N-nitroso-duloxetine impurity above safety assessment limit

    FDA · DrugBreckenridge Pharmaceutical, Inc.U.S NationwideOfficial notice →
  5. 2025-08-06
  6. US
    Drugs & medicalClass II

    3M Company recalls 3M Ranger Irrigation Fluid Warming Set, REF 24750

    Notice was issued to clarify flow rates related to the inlet fluid temperature.

    FDA · Device3M CompanyUS (nationwide) and Argentina, Australia, Austria, Belgium, Bosnia and…Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

    Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride,…

    FDA · DeviceSiemens Healthcare Diagnostics Incworldwide distribution - US Nationwide and the countries of Canada, Chile,…Official notice →
  8. US
    Drugs & medicalClass II

    Remel, Inc recalls Blood Agar,5% Sheep Blood 100/PK, Product Number R01202

    The products may contain surface and subsurface contamination of Listeria monocytogenes.

    FDA · DeviceRemel, IncUS Nationwide distribution.Official notice →